Boonton 4540 Series RF Power Meter
Remote Operation
5-7
*ESR?
Description:
Return the current value of the Standard Event Status Register, then clear the register. This register
has bits assigned to a number of possible events or conditions of the instrument. When the event
occurs, the corresponding bit is latched. The register value is read using this command.
Individual bits may be enabled or disabled for SRQ generation using the ESE mask (see *ESE
command). The following table shows the bit assignments in the Standard Event Status Register:
Bit Value Definition
0
1
Operation Complete Flag 1 = all current operations have completed execution.
1
2
Not used
2
4
Not used
3
8
Device Dependent Error 1 = the instrument encountered a device dependent error.
4
16
Not used
always returns 0
5
32
Command Error
1 = a remote interface command error exists.
6
64
Not used
always returns 0
7
128
Not used
always returns 0
Syntax:
*ESR?
Returns:
Current Value of Event Status Register (0 to 255)
Valid Modes:
Any
*IDN?
Description:
Return the instrument identification string. This string contains the manufacturer, model number,
serial number and firmware version number.
Syntax:
*IDN?
Returns:
< Mfgr, Model#, Serial#, Version# >
Valid Modes:
Any
*OPC
Description:
Clears the OPC (Operation Complete) status flag. This command is issued before the command to
be checked for completion. After this, the flag may be queried by *OPC? until a value of one is
returned, indicating the command has completed. Note that the query is not a true query - a value
of zero will never be returned.
Syntax:
*OPC
Argument:
None
Valid Modes:
Any
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...