![Boonton 4541 Скачать руководство пользователя страница 167](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863167.webp)
Boonton 4540 Series RF Power Meter
Remote Operation
5-33
MARKer:POSItion:TIMe
Description:
Set or return the time (x-axis-position) of the selected marker relative to the trigger. Note that
time markers must be positioned within the time limits of the trace window in the graph display. If
a time outside of the display limits is entered, the marker will be placed at the first or last time
position as appropriate.
Syntax:
MARKer[1|2]:POSItion:TIMe <numeric_value>
Argument:
<numeric_value> = display-time in seconds (see restrictions)
Valid Modes:
Pulse and Modulated Modes.
Restrictions:
For zero delay trigger position in the center of the display , the following relationship must be
satisfied: TrigDly - (5*time/div) < MarkerTime < T (5*time/div) where the timebase
setting is time/div