![Boonton 4541 Скачать руководство пользователя страница 151](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863151.webp)
Boonton 4540 Series RF Power Meter
Remote Operation
5-17
CALibration[:INTernal]:AUTOcal
Description:
Performs a multi-point sensor gain calibration of the selected sensor with the internal 50 MHz
calibrator. This procedure calibrates the sensor’s linearity at a number of points across its entire
dynamic range. The calibration source defaults to the instrument’s internal calibrator if not
specified.
Syntax:
CALibration[1|2][:INTernal]:AUTOcal[?]
Returns:
0 if successful, 1 otherwise (using query form only)
Valid Modes:
Peak and CW Power sensors only
CALibration[:INTernal]:FIXedcal
Description:
Performs a calibration at a fixed frequency and level. Fixed-cal does not provide for automatic
control of the internal 50 MHz calibrator. The RF output level of the calibrator in use must be set
by the user. Fixed-cal assumes that a valid Zero has already been performed.
Syntax:
CALibration[1|2][:INTernal]:FIXedcal[?]
Returns:
0 if successful, 1 otherwise (using query form only)
Valid Modes:
CW Power sensors only
CALibration[:INTernal]:ZERO
Description:
For a sensor that has been previously auto calibrated and has valid auto-cal status, performs a zero
offset null adjustment. The sensor does not need to be connected to any calibrator for zeroing –
the procedure is often performed in-system. However, this command will turn off the internal
calibrator prior to zeroing to avoid the need to perform this step explicitly.
Syntax:
CALibration[1|2][:INTernal]:ZERO[?]
Returns:
0 if successful, 1 otherwise (using query form only)
Valid Modes:
Any
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...