Boonton 4540 Series RF Power Meter
Remote Operation
5-15
CALCulate:STATe
Description:
Set or return the measurement state of the selected channel. When ON, the channel performs
measurements; when OFF, the channel is disabled and no measurements are performed.
Syntax:
CALCulate[1|2]:STATe <Boolean>
Argument:
<Boolean> = { 0, 1, OFF, ON }
Valid Modes:
Any
CALCulate:UNITs
Description:
Set or return units for the selected channel. For power sensors, voltage is calculated with
reference to the sensor input impedance. Note that for ratiometric results, logarithmic units will
always return dBr (dB relative) while linear units return percent.
Syntax:
CALCulate[1|2]:UNITs <character data>
Argument:
<character data> = {DBMw, Watts, Volts, DBV, DBMV, DBUV}
Valid Modes:
Any