4.5.46 System > Calibrator Menu (External 1 GHz - optional)
The Calibrator Menu is used to seclect and control the external 1 GHz RF Calibrator (Model 2530). It can also select the 50
MHz internal calibrator. The external calibrator may be pulse modulated using either a built-in pulse generator, or via a rear-
panel BNC pulse input. Note that this menu does not contain any items related to sensor calibration—it is only for controlling
the calibrator for use as a signal source. For sensor calibration information, refer to Section 3.5 and Section 4.4.2 of this
manual.
Main Pg2 > System > Calibrator > Cal Output
Description:
Enable or disable the RF output of the 1 GHz external
calibrator and select the mode of operation.
Menu Type:
Multiple Choice.
Choices:
Off: Disabled
CW: Provides 1 GHz CW output
Int Pulse: Provides pulse modulation from internal
pulse generator.
Ext Pulse: Provides pulse modulation from external
signal to BNC input connector.
SCPI Command:
OUTPut:EXTernal:SIGNal <Boolean>
Main Pg 2 > System > Calibrator > Set Level
Description:
Set the output of the 1 GHz external calibrator in dBm.
Menu Type:
Numeric value.
Range:
-60.0 dBm to +20.0 dbm (Internal 50 MHz)
-50.0 dBm to +20.0 dbm (External 1 GHz)
SCPI Command:
OUTPut:EXTernal:LEVel:[POWer]
<numeric_value>
Main Pg 2 > System > Calibrator > Sel Calibrator
Description:
Select the active calibrator for calibration and control. ―Int
‖
selects the internal 50 MHz
calibrator. ―Ex
t‖
selects an external 1 GHz calibrator. "Ext" cannot be selected if an
external calibrator is not connected. An error message pop-up will appear.
Menu Type:
Toggle.
Choices:
Int, Ext
SCPI Command:
None.
Fig.4-58 Calibrator Menu
External 1 GHz
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...