Boonton 4540 Series RF Power Meter
Remote Operation
5-43
OUTPut:MIO:STATus:SETTing
Description:
Set/return Multi IO status out setting. Below is a description of each argument:
Argument
Description
ALMACT:
Sets the Multi IO output to +5 VDC when an alarm is currently active.
Multi IO output will be 0 VDC if no alarms are active.
ALMLATCH:
Sets the Multi IO output to +5 VDC when an alarm limit has been
exceeded. Multi IO output will be 0 VDC if no alarm limits have been
exceeded.
MEASRDY:
Multi IO output is 5 VDC when a measurement is ready to be read.
CAL:
Multi IO output is 5 VDC when a calibration is complete.
Syntax:
OUTPut:MIO:STATus:SETTing <character data>
Argument:
<character data> = {
ALMACT, ALMLATCH, MEASRDY, CAL
}
Valid Modes:
Any
OUTPut:MIO:STATus:SOURce
Description:
Set/return Multi IO source for status out (alarm). Single channel units (Model 4541) only accept
―CH1
‖.
Syntax:
OUTPut:MIO:STATus:SOURce <character data>
Argument:
<character data> = {
CH1, CH2, BOTH
}
Valid Modes:
Any
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...