![Boonton 4541 Скачать руководство пользователя страница 176](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863176.webp)
Boonton 4540 Series RF Power Meter
Remote Operation
5-42
OUTPut:[MIO:]RECOroderMIN
Description:
Set or return the Multi IO recorder output minimum, or downscale (-10.0V or 0.0V) power
reference level. For voltage probes, this is the equivalent power at the current impedance setting.
Syntax:
OUTPut:RECOrder:MIN < numeric_value >
Argument:
< numeric_value > = -200.00 to +200.00 dBm
Valid Modes:
Any
OUTPut:[MIO:]RECOroder:
POLarity
Description:
Set or return the recorder output polarity. UNIPOLAR selects 0.0 volts and 10.0 volts to as
minimum and maximum output levels, and BIPOLAR selects -10.0 to +10.0 volts.
Syntax:
OUTPut:RECOrder:POLarity <character data>
Argument:
<character data> = UNIPOLAR, BIPOLAR
Valid Modes:
Any
OUTPut:[MIO:]RECOroder:
SCALing
Description:
Set or return the Multi IO measurement mode when in recorder mode. If set to AUTO, the output
level is automatically scaled to match the display, and generally will ―downsca
le‖
every decade as
power increases. In MANUAL mode, the output level is scaled using the minimum and maximum
powers set with OUTPut:[MIO:]RECOrder:MIN and :MAX as the downscale and full scale
values. This allows very wide range or very high resolution.
Syntax:
OUTPut:[MIO:]RECOrder:SCAling <character data>
Argument:
<character data> = AUTO, MANUAL
Valid Modes:
All
OUTPut:[MIO:]RECOroder:
SOURce
Description:
Set or return the source channel for the Multi IO recorder output.
Syntax:
OUTPut:[MIO:]RECOrder:SOURce <character data>
Argument:
<character data> = CH1, CH2
Valid Modes:
All
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...