![Boonton 4541 Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863020.webp)
Boonton 4540 Series RF Power Meter
General Information
1-6
1.6 Specifications
Performance specifications for the 4540 Series are listed in Table 1-2.
Performance specifications for all Boonton power sensors are found in the
Boonton Sensor Manual,
which
may be ordered as Boonton p/n 98501900J.
Table 1-2 4540 Series Performance Specifications
(Specifications are subject to change without notice)
SENSOR INPUTS
RF Frequency Range:
1 MHz to 110 GHz*
Pulse Measurement Range:
-50 to +20 dBm*
Modulated Measurement Range: -55 to +20 dBm*
CW Measurement Range:
-70 to +44 dBm*
Relative Offset Range:
±300.00 dB
Vertical Scale: Logarithmic
0.1 to 50 dBm/div in 1-2-5 sequence
0.1 to 50 dBV/div in 1-2-5 sequence
0.1 to 50 dBmV/div in 1-2-5 sequence
0.1 to 50 dBµV/div in 1-2-5 sequence
Linear
1 nW/div to 50 MW/div in 1-2-5 sequence
1 nV/div to 50 MV/div in 1-2-5 sequence
Rise time/Video Bandwidth:
7 ns / 70 MHz*
Single-Shot Bandwidth:
5 MHz (based on 10 samples per pulse)
Pulse Repetition Rate:
30 MHz max (for internally triggered measurements)*
Minimum Pulse Width:
15 ns*
* SENSOR DEPENDENT
MEASUREMENT SYSTEM
Sensor inputs:
One or two sensor measurement channels.
Measurement Technique:
Random repetitive sampling system that provides pre and post-trigger
data as well as statistical histogram accumulation.
Maximum Sampling Rate:
50 Mega-samples/second on two channels simultaneously. (Equivalent
effective sampling rate of 5 Giga-samples/second)
Memory depth:
262,144 samples at max sampling rate
Vertical Resolution:
0.008%, 14-bit A/D Converter
TIME BASE
Time Base Range:
10 ns/div to 1 hr/div
Time Base Accuracy:
0.01%
Time Base Resolution:
0.2 ns
Time Base Display:
Sweeping or Roll Mode
Trigger Delay Range:
10 ns - 500 ns timebases -4 ms to +100 ms
1 µs - 10 ms timebases
4000 divisions
20 ms - 3600s timebases -40 to +100 s
Trigger Delay Resolution:
0.02 divisions
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...