SENSe:ARRay:CALTemp?
Description:
Returns both channels’ sensor calibration temperatures.
Syntax:
SENSe:ARRay:CALTemp?
Returns:
CC1, Sensor1 cal temp in degrees C, CC2, Sensor2 cal temp in degrees C
Where CCn is the measurement condition code.
Valid Modes:
All
SENSe:CALTemp?
Description:
Returns the selected channel’s sensor calibration temperature.
Syntax:
SENSe[1|2]:CALTemp?
Returns:
CC, Sensor[1|2] cal temp in degrees C
Where CCn is the measurement condition code.
Valid Modes:
All
SENSe:CORRection:CALFactor
Description:
Set or return the frequency calfactor currently in use on the selected channel. Note setting a
calfactor with this command will override the ―autom
atic‖
frequency calfactor that was calculated
and applied when the operating frequency was set, and setting the operating frequency will
override this calfactor setting.
Syntax:
SENSe[1|2]:CORRection:CALFactor <numeric_value>
Argument:
<numeric_value> = –3.00 to 3.00 dB
Valid Modes:
All
SENSe:CORRection:DCYCle
Description:
Set or return the duty cycle correction factor currently in use on the selected channel.
Syntax:
SENSe[1|2]:CORRection:DCYCle
Argument:
<numeric_value> = 0.01 to 100.00 percent
Valid Modes:
CW Sensor (Modulated Mode)
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...