Boonton 4540 Series RF Power Meter
Remote Operation
5-64
STATus:QUEStionable:CALibration:CONDition?
Description:
Return the current value of the Questionable Calibration Condition register. This register is used
to notify the user of questionable quality with respect to calibration. The following table shows
the bit assignments in the register. These bits are updated by the instrument in real time, and can
change in response to changes in the instrument’s operating condition.
Bit
Value
Definition
0
1
Sens1 Needs Cal
1 = Channel 1 sensor needs calibration.
1
2
Sens2 Needs Cal
1 = Channel 2 sensor needs calibration.
2
4
Sens1 Default Shape
1 = Channel 1using default shape table.
3
8
Sens1 Default Shape
1 = Channel 2 using default shape table.
Syntax:
STATus:QUEStionable:CALibration:CONDition?
Returns:
16-bit register value (0 to 32767)
Valid Modes:
Any
STATus:QUEStionable:CALibration:ENABle
Description:
Sets or returns the Questionable Calibration Enable register, which contains the bit mask that
defines which true conditions in the Questionable Calibration Event register will be reported in the
Questionable Calibration Summary bit of the Questionable Condition register. If any bit is 1 in
the Questionable Calibration Enable register and its corresponding Questionable Calibration Event
bit is true, the Questionable Calibration summary bit will be set.
Syntax:
STATus:QUEStionable:CALibration:ENABle <numeric_value>
Argument:
<numeric_value> = 0 to 32767
Valid Modes:
Any
STATus:QUEStionable:CALibration:EVENt?
Description:
Returns the current contents of the Questionable Calibration Event register then resets the register
value to 0. The Questionable Calibration Event register contains the latched events from the
Questionable Calibration Condition register as specified by the Questionable Calibration status
group’s positive and negative transition filters.
Syntax:
STATus:QUEStionable:CALibration:EVENt?
Returns:
16-bit register value (0 to 32767)
Valid Modes:
Any
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...