![Boonton 4541 Скачать руководство пользователя страница 120](http://html.mh-extra.com/html/boonton/4541/4541_instruction-manual_2787863120.webp)
Boonton 4540 Series RF Power Meter
Operation
4-72
4.5.47 System > Calibrator > Pulse - Preset Menu (Ext 1 GHz - opt.)
The pulse menu is specific to the 1 GHz calibrator when it is connected. Control of the internal pulse generator is availble in
2 forms. Preset mode provides a limited choice of periods and duty-cycles, but is easy to use. The Variable mode provides a
wide choice of periods and pulse widths, but with some restrictions.
Main Pg 2 > System > Calibrator > Pulse > Polarity
Description:
Selects the pulse polarity.
Menu Type:
Toggle
Choices:
+, -
SCPI Command:
None
Main Pg 2 > System > Calibrator > Pulse > Duty Cycle
Description:
Selects the pulse duty cycle in percent.
Menu Type:
Popup Selection
Choices:
10%, 20%, 30%, 40%, 50%
60%, 70%, 80%, 90%
SCPI Command:
OUTPut:EXTernal:PULSe:DCYCle <numeric_value>
Main Pg 2 > System > Calibrator > Pulse > Period
Description:
Selects the pulse period.
Menu Type:
Multiple Choice
Choices:
10 msec, 1 msec, 100 usec
SCPI Command:
OUTPut:EXTernal:PULSe:PERiod <numeric_value>
Main Pg 2 > System > Calibrator > Pulse > Pulse Control
Description:
Selects the pulse control method.
Menu Type:
Toggle.
Choices:
Preset, Variable
SCPI Command:
OUTPut:EXTernal:PULSe:CTRL
<character data>
Fig. 4-60 External 1 GHz
Calibrator Menu - Pulse
Preset
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...