Step 7: Insert Bloom Test strip into Bloom Lab
1. Insert the Bloom Test strip into the Bloom Lab . You
should feel the Bloom Test strip lock into place. The
Bloom Lab will show a white heartbeat pattern on the
LED ring when the test strip is inserted correctly.
Insert Bloom Test strip into the Bloom Lab.
2. Processing begins. The progress of the measurement is
communicated by the LED ring of the Bloom Lab with
a spinning white loader. The Bloom App shows the
estimated time remaining.
3. Once processing has been completed, the Bloom App
indicates the successful completion of the testing.
Caution:
• Place the Bloom Lab on a stable, fl at surface and do not
move the device during measurement.
• Do not remove the Bloom Test strip from the Bloom Lab
during the testing procedure.
• Ensure that the Bloom Lab does not come into contact
with blood.
• Make sure that the Bloom Test strip is clean and free
from dust before inserting it into the Bloom Lab.
Possible errors:
• In case of a recoverable error, the LED ring of the Bloom
Lab will show a red pattern, indicating that something
went wrong. See section 7.1. for further details.
• In case of a fatal error, such as a general error requiring
service of the Bloom Lab or power adapter, the LED ring
of the Bloom Lab will show a solid red ring, indicating that
a general hardware failure error occurred.
Step 8: Dispose components
The used Bloom Thyroid Test components are potentially
biohazardous. Please dispose of them carefully according to
applicable laws and regulations.
4.3. Process sampling and report collection
The final step of the testing process is to complete the
questions needed to compile the personalized Bloom Thyroid
Report using the Bloom App. While the Bloom Lab is process-
ing the test result, the Bloom App will prompt you to do the
following:
Check the Bloom App
1. Answer the questions. When you are presented with a
multiple-choice question, choose all answers that apply
and then confirm to proceed.
2. After around 20 minutes, the results from the Bloom
Lab will be transmitted to the Bloom App. The app will
alert you that the Bloom Thyroid Report is ready.
3. View the personalized Bloom Thyroid Report
Information: If the Bloom App disconnects from the Bloom
Lab during the measurement process, the Bloom Lab com-
pletes the measurement process and saves the measure-
ment value locally for a 3-hour period. If the same Bloom
App reconnects with the Bloom Lab during this period, the
saved measurement value will be sent to the Bloom App
and the Bloom Lab will subsequently delete the measure-
ment value and any unique identifi er of the Bloom App.
After a 10-minute period other Bloom Apps will be able
to connect to the Bloom Lab and perform new tests, which
will automatically delete the previous results.
4.4. Device cleaning
Caution: Please use protective equipment, such as gloves,
when cleaning the surfaces of the Bloom Lab to avoid
contamination.
After performing the Bloom Thyroid Test, please make sure
to leave a clean workspace. Thoroughly wipe all surfaces of
the Bloom Lab after every use with the provided disinfec-
tion cleaning wipe. Please check gaps for stains and clean
thoroughly when needed.
5. Storage
In order to avoid hazardous situations, the Bloom Thyroid
Test should be stored with care at all times, this includes:
• The recommended storage temperature b8°C
and +30°C (46°F and 86°F)
• A dry indoors environment without direct or excessive heat
and humidity below 80% (bathrooms not suitable)
• Out of reach for children to avoid injury through misuse of
components like lancets.