background image

 
 

V 1.31 

 

2021. 05 

48

  / 

171

 

Doc. # : BN-OP-C7 

When the user selects the line of ID, NAME, Department, a keyboard will show up on the 

screen. The default of every line is a blank and if there is a certain figure on the line, it 
should be shown exactly when loading the keyboard window. 

When the user wants to set the data after inputting, the user can press OK button on the 

keyboard and if the user wants to cancel the setting, pressing ESC button will do it and the 
keyboard window will disappear. 

 

 

 

NOTE 

-

 

When entering an ID, it is impossible to 

enter special characters such as “, <, >, ?, /, 

*, |, :, \, etc. (these invalid characters are set to be inactive) 

-

 

When entering an ID, an English keyboard will be displayed, regardless of whether 
there is a multi-language set-up. 

-

 

When  entering  an  ID  by  connecting  to  an  external  keyboard,  proceed  by  using 
general  alphabet  and  numbers  only.  If  you  enter  an  ID  using  Latin  extended 
characters and Russian, an error may occur when transferring files via PC and USB. 

Содержание Cardio7

Страница 1: ...Copyright 2021 By Bionet CO LTD All rights reserved Cardio7e Cardio7 ECG Operation Manual Ver 1 31 2021 05 31 www ebionet com...

Страница 2: ...r 1 year 2 years in EU But the warranty for accessories runs for 6 months This product is guaranteed against faulty workmanship for a period of 1 year Under this warranty we undertake to repair free o...

Страница 3: ...61 Digital ro 31 gil Guro gu SEOUL 08375 REPUBLIC OF KOREA Tel 82 2 6300 6410 Fax 82 2 6499 7789 E mail sales ebionet com Website www ebionet com U S A sales service representative Bionet America Inc...

Страница 4: ...due to customer s unskilled installation For offering service when a foreign substance is introduced or mis cleaned by the customer Fee charged 1 The warranty does not cover cleaning adjusting and in...

Страница 5: ...service agents are not responsible for damage resulting from inappropriate use or carelessness WARNING There may be serious injuries fatal accidents or financial damages if you violate this instructio...

Страница 6: ...ambient temperature 10 to 40 50 to 104 and humidity 30 to 85 Near electronic heaters High humidity or ill ventilated place Place where excessive shock or vibration may occur Place exposed to flammable...

Страница 7: ...product near a flammable anesthetic or solvent Continuous operation IEC EN60601 1 2 Electromagnetic Compatibility Requirements standard Classification Description Class A The device or system is suit...

Страница 8: ...electrodes is required to ensure successful defibrillation WARNING ELECTRODES Polarizing electrodes stainless steel or silver constructed may cause the electrodes to retain a residual charge after de...

Страница 9: ...at they may present a stumbling hazard For safety reasons all connectors for patient cables and lead wires are designed to prevent inadvertent disconnection should someone pull on them For devices ins...

Страница 10: ...ified by the manufacturer CAUTION ELECTRIC SHOCK To reduce the risk of electric shock do not remove cover or back Refer servicing to qualified personnel CAUTION OPERATOR Medical technical equipment su...

Страница 11: ...use of the device Safety Sign To signify that the instruction manual must be read Reading the instruction manual before starting work or before operating equipment General prohibition sign Defibrillat...

Страница 12: ...AC Power Connection Indicator Manufacturer name and address Authorized European representative Waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must b...

Страница 13: ...rt 34 Chapter 2 Preparation for Electrocardiography 40 1 Location of Electrodes 41 2 Electrode Connection 42 3 Starting ECG recording 44 4 Basic Setup 45 5 LCD Monitor display 56 6 Disclosure 57 7 Rhy...

Страница 14: ...klist management 149 1 Display and function 150 2 ECG Test 151 3 Data deleting 152 4 System Setup 153 5 Switching menu 154 6 Patient information 154 7 Data search 155 Chapter 5 System Management 156 1...

Страница 15: ...CGs During Defibrillation 4 Product Features 5 Product Configuration Basic Components and Accessories Optional components Body Configuration Front Panel Graphic display LCD Control Panel Power 6 Syste...

Страница 16: ...ion for use The Cardio7e Cardio7 ECG Analysis System is intended to acquire analyze display and record electrocardiographic information from adult and pediatric populations Basic systems deliver 12 le...

Страница 17: ...y of the electrode to allow the ECG trace to return after defibrillation We recommend using non polarizing disposable electrodes with defibrillation recovery ratings as specified in AAMI EC12 4 2 2 4...

Страница 18: ...check for any damage to the body and accessories Basic Components and Accessories Cardio7e Cardio7 body 1 EA Dimension 300 W x 290 D x 89 5 H mm Patient cable 1 EA Length 3 700mm Max Limb electrodes...

Страница 19: ...when you use nonstandard or other brand accessories We strongly recommend you use only the authorized accessories which we supply WARNING How to replace the battery If the battery provided by us is n...

Страница 20: ...V 1 31 2021 05 20 171 Doc BN OP C7 Body Configuration Top View Handle Printer Cover Printer Cover Switch LCD Control Panel...

Страница 21: ...V 1 31 2021 05 21 171 Doc BN OP C7 Front View Printer Cover Switch...

Страница 22: ...rotective Ground Terminal Power Switch AC Power Connection Port Fuse 250V 3 15AL x 2 USB Port RJ45 LAN Port NOTE Appliance inlet of separate power supply unit is used as mains disconnecting device NOT...

Страница 23: ...risk of electric shock if the Rest stand of the equipment is damaged or can not be fixed to the product Do not use the product and immediately ask the manufacturer and the seller for repair NOTE To av...

Страница 24: ...V 1 31 2021 05 24 171 Doc BN OP C7 Front Panel During device boot up you can see the system software version and the Bionet name Rotary key Power status Battery status Button Graphic display window...

Страница 25: ...eart Rate Displays Patient ID Displays Battery Status or AC Power Connection Status Displays ECG Sensitivity with One of 5 10 20 and Auto mm mV Displays Printing Speed with One Of 5 12 5 25 50 And 100...

Страница 26: ...are saved and are printed as an output form after processing Process the recorded data with or without revision and print them out It operates the product in the most frequently used operation from th...

Страница 27: ...nt must be connected to a power supply with a protective earth Battery Power The device will get its power from the battery and the unit will display the battery power icon shown above when the system...

Страница 28: ...ng battery use will reduce battery life and increase the frequency of replacement Recharge the battery after it is discharged to prevent pre discharge Display Battery Power Status Battery Fully Charge...

Страница 29: ...properties it is self discharged The self discharge rate increases by twice every 10 C 18 F increase in temperature Battery retention is more lost in the higher temperatures Depending on battery life...

Страница 30: ...erative product Remove the battery from the system and put it in a separate bin WARNING Don t keep at high temperatures or burned of a battery due to there is a risk of explosion You can wear serious...

Страница 31: ...t or flammable materials Power Connection The equipment needs electrical power to operate Plug in one end of the power cable to wall socket and the other to Cardio7e Cardio7 Patient Cable Connection C...

Страница 32: ...ing to this user guide and network specifications LAN connection specifications are described in the following standard specifications Wired network IEEE 802 3 Wireless network IEEE 802 11 a b g n If...

Страница 33: ...ted Data is transmitted in an incomplete state or not transmitted at all Data may be transmitted to the wrong server Data may be blocked forged or damaged in case there is no independent network envir...

Страница 34: ...name will be displayed when the switch is turned on and then after approx 10 seconds the initial menu selection will appear on the screen as shown in the figure below You can move to the main page of...

Страница 35: ...V 1 31 2021 05 35 171 Doc BN OP C7 The following is the main page of ECG menu If you remember the information shown below it will be easier for you to use the main page of ECG...

Страница 36: ...tudy Queue screen by touching the icon USB memory connection mark Network status mark wired wireless Move to the Network Setup screen by touching the icon The error icon will appear in the event of an...

Страница 37: ...page of ECG file management Menu Description Going to Patient List To skip to previous page To skip to next page To check the patient s information To search the data by inputting what you want to fin...

Страница 38: ...e of ECG worklist management Menu Description Going to Patient List To skip to previous page To skip to next page To check the patient s information To search the data by inputting what you want to fi...

Страница 39: ...an error occurs during transfer to the PACS server Memorizing the following at the Study Queue Management screen will help more convenient use NOTE This icon appears on the top right side when data t...

Страница 40: ...ermeasures for Bad Lead Connection 3 Starting ECG recording 4 Basic Setup General Patient Information Input Sensitivity Setup Printing speed Setup 5 LCD Monitor display 6 Disclosure 7 Rhythm Mode Prin...

Страница 41: ...LL F Left Leg RA R Right Arm LA L Left Arm Chest electrodes are located as follows V1 C1 Fourth Intercostal Space at The Right Border of The Sternum V2 C2 Fourth Intercostal Space at The Left Border...

Страница 42: ...you can connect electrodes for the chest to the terminal of V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 and V6 C6 How to Attach Electrodes Lay down the patient on the bed and let the patient relax Attach the elect...

Страница 43: ...d will NOT be displayed when set to OFF Please conduct an ECG test with the lead connection is correct The following two cases could cause a lead fault error First case leads are detached from the pat...

Страница 44: ...s the COPY button to print out the previously saved 10 second data or modified data Press the RHYTHM key to monitor the ECG waveform through the printer not through the LCD Press the ESC key to stop t...

Страница 45: ...tary key or touch screen With the rotary key the desired menu and changed configuration values can be set easily Same functionality can also be achieved with the touch screen by pressing and changing...

Страница 46: ...lect values in the same manner as menu movement shown above When you select either 5mm sec 12 5mm sec 25mm sec 50mm sec or 50mm sec at the Speed Setting POP UP the POP UP window will disappear Using t...

Страница 47: ...7 Patient Information Input In this menu operators can input the patient s ID name age sex height weight race and smoking habit NOTE The patient s surname is available after changing the file version...

Страница 48: ...user wants to cancel the setting pressing ESC button will do it and the keyboard window will disappear NOTE When entering an ID it is impossible to enter special characters such as etc these invalid c...

Страница 49: ...t the year of birth for 1 year old enter age in the form of week or day for 1 year old Automatic diagnosis is available from postnatal day 1 for pediatrics When the user wants to set the data after in...

Страница 50: ...or Female will be shown respectively After selecting the data which you want input all the information of the patient When pressing OK button the information will be saved while pressing cancel butto...

Страница 51: ...of White Asian Black will show up After selecting the data which you want to input all the information of the patient When pressing OK button the information will be saved while pressing cancel button...

Страница 52: ...information you will see the window as shown below After checking all information then pressing OK button the information will be saved while pressing cancel button will cancel the setting When press...

Страница 53: ...rers implementation of input methods you need to verify the scanner whether it is supported by Bionet Input methods supported by Bionet International standards USB Products below are tested and confir...

Страница 54: ...matically set the sensitivity level for each channel for example limb electrodes I II III aVR aVL aVF to 10 mm mV and chest electrodes V1 V2 V3 V4 V5 V6 to 5mm mV Note that 10mm mV means printing 1mV...

Страница 55: ...m sec width is equal to half of the 25 mm sec width the 5 mm sec with is equal to 1 5 of the 25 mm sec width The 50mm sec width is equal to twice that of the 25mm sec width the 100mm sec width is equa...

Страница 56: ...her the signals from all channels are coming out properly the other one is to see the status of the electrocardiogram of the patient for an extended time The printing speed can be selected to 5mm s 12...

Страница 57: ...here the user can select it to perform diagnosis output and data transfer The 10 second square field on the Disclosure Window can be moved in 1 second increments by using rotary right left keys front...

Страница 58: ...and signal level can be set to 2 5mm mV 5mm mV 10mm mV 20mm mV and Auto When pressing the RHYTHM key on the control panel you will see the message as shown below and printing will start in accordance...

Страница 59: ...form that are set up by the user 10 seconds ECG recording When one of 3ch 1rhy 6ch 1rhy 12ch 3ch 3rhy and beat is selected for the channel form press the RECORD key to display the saving time as belo...

Страница 60: ...the bottom 6CH 1 ST I II III aVR aVL and aVF will be recorded for the first 4 seconds and V1 V2 V3 V4 V5 and V6 for the next 4 seconds And 1 channel of rhythm will be recorded for 10 seconds at the bo...

Страница 61: ...V 1 31 2021 05 61 171 Doc BN OP C7 Rhythm Report Form 3CH Rhythm Report Form 6CH...

Страница 62: ...V 1 31 2021 05 62 171 Doc BN OP C7 Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...

Страница 63: ...V 1 31 2021 05 63 171 Doc BN OP C7 Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...

Страница 64: ...V 1 31 2021 05 64 171 Doc BN OP C7 Record Report Form 12 CH Record Report Form 6 CH 1 RHY ST Map...

Страница 65: ...V 1 31 2021 05 65 171 Doc BN OP C7 Record Report Form BEAT REPORT TEXT Record Report Form BEAT REPORT VECTOR...

Страница 66: ...V 1 31 2021 05 66 171 Doc BN OP C7 Record Report Form BEAT REPORT GUIDE Record Report Form BEAT REPORT ST Map Dot...

Страница 67: ...V 1 31 2021 05 67 171 Doc BN OP C7 Record Report Form BEAT REPORT ST Map Bar...

Страница 68: ...RA R Amplitude RD R Duration SA S Amplitude SD S Duration QRSD QRS Duration QRSA QRS Amplitude TA T Amplitude STA ST Amplitude QTc collect Q T Interval ST60A ST60ms Amplitude ST80A ST80ms Amplitude N...

Страница 69: ...f P QRS and T are all reversed on Lead I aVR and aVL as well as lead II and III are switched with each other As you proceed from V1 to V6 R wave is reducing on Chest lead For patients with dextrocardi...

Страница 70: ...ssing the COPY key following an EKG recording After pressing the COPY Key the following message will be displayed on the LCD screen and the printing process will start If Copy Mode Printing is called...

Страница 71: ...ill start printing after showing the message as below When printing is abnormally completed the state of the system setting will be displayed after showing the message as shown below for a second If y...

Страница 72: ...etup Network Setup Hospital Setup ECG Setup Record Setup and Service Setup and you can change the setting at each window The main page shows Basic Setup After finishing the settings press OK to save t...

Страница 73: ...Doc BN OP C7 Basic Setup When choosing Basic menu at the left side of the Setup page you will see the page for Basic Setup Basic Setup consists of Date Time Touch screen Start Option Unit Date Type a...

Страница 74: ...ic button then the focus will go to the Date item By using the touch screen or rotating the rotary switch choose the Date yyyy mm dd or Time hr min sec then the keypad will show up You can set up the...

Страница 75: ...for setting calibration will appear as shown below Follow the instructions shown on the screen and then the coordinates will be set up Please input the information in the right place otherwise the to...

Страница 76: ...nian Russian Spanish and Turkish Select the language you want to use and click the OK button and then you can use the service in the chosen language Note Patient information previously entered will be...

Страница 77: ...can set up the index page shown at the time of initial boot up according to the common dialog When clicking the Start Option button a small menu window will appear and you can select the initial page...

Страница 78: ...7 Units This menu is to select the units for height and weight of the patient When you click Units button cm kg or Inch Lbs will be shown If you select Inch Lbs unit the patient s height information w...

Страница 79: ...e OK button the date format is converted to the one selected and can be seen on the output Device Name Setup Device Name refers to the name of the device The user can input the name of the device in o...

Страница 80: ...ected as Wired or Wireless in Net Device setup The automatic IP DHCP or fixed Manually can be selected through the Configure If the Configure is setup as automatic IP DHCP the Device IP Subnet Mask Ga...

Страница 81: ...the equipment in the Connect to menu After selecting the type of interlocking server select the Setting button an appropriate setup window to the choice of server will be displayed Note When setting u...

Страница 82: ...led PC will be shown on the input box Choose the IP of the PC you intend to connect as the IP of a BMS Server installed PC If multiple IPs are displayed choose one belonging to the PC to connect At th...

Страница 83: ...ing Ping button Ping 100 Succeeded message will be shown for a successful connection whereas Ping Failed message will be shown for connection failure Note Wired usage is recommended for a BMS Server i...

Страница 84: ...for related settings such as the AE Title IP and Port of Cardio7e Cardio7 Worklist Server and Store Server will appear When the input window of each item is selected the keyboard window will appear En...

Страница 85: ...e Verify Success message will appear when it has failed Verify C Echo Failed will appear Set the PACS option by pressing the More button when connection to the server is confirmed Spiro Exam Code Ente...

Страница 86: ...o Return Worklist data will be transferred to the PACS Server and the Worklist screen will re appear automatically by using the Exam button after selecting the patient to measure or by using the NETWO...

Страница 87: ...PDF or RAW after checking the format supported by the PACS Server Retry Count Select the number of Retransmission attempts when a communication error occurs during transfer of data to the PACS Server...

Страница 88: ...should be selected according to the character set in order to present contents such as patient information by language when transferring a DICOM file to the PACS Server Upon the completion of each se...

Страница 89: ...settings such as Work type GDT Directory Component name File name and Image type Work type Set how the GDT function of the system works Server Cardio7e Cardio7 receives requests and commands Client C...

Страница 90: ...e Select the name form of the file to be shared by the GDT protocol Image Select the image format of the data file to share with the GDT protocol When it selects the Input window for each item a keybo...

Страница 91: ...ing menu The screen for related settings such as DNS Server URL ID PW Path appears DNS Server The IP of the server that supports the domain name system URL Web address to access ID User ID to be used...

Страница 92: ...ach item a keyboard window appears and you can enter the content After entering you can check the connection between Worklist Server and Store Server by pressing the Verify button The message Verify S...

Страница 93: ...ork and Set the Device as Wireless When selection is done properly the Survey button on the bottom will be activated 3 Select the Search button to search for the AP The AP is to be searched automatica...

Страница 94: ...t time the security key input window will appear if the Encrypt field is On Note The user should set the AP Search and Security Key in order to use the wireless network If there is no security set on...

Страница 95: ...pted and the message describing the connection status will appear 7 When the message reporting a proper connection appears click the Close button When connection is successful the wireless icon on the...

Страница 96: ...ake or illegible characters can appear Note It is recommended connecting up to eight the same network under normal AP status Due to the nature of wireless connectivity may be poor depending on the env...

Страница 97: ...V 1 31 2021 05 97 171 Doc BN OP C7 ECG Setup When selecting the ECG menu at the left side of the Setup page ECG Setup page will show up at the right side...

Страница 98: ...H printing To print 3 channels at the same time I III aVR aVF V1 V3 V4 V6 If you want to change the channel press RHYTHM on the control panel 6CH printing To print 6channels at the same time I aVF V1...

Страница 99: ...Report will be selected and when clicking it again Continue will be selected Default is set as Continue Report To print out 10 second data in the form of Report Continue The leads will be printed out...

Страница 100: ...ard ECG recording paper instead of general fax paper thermal paper undo the grid marking setup as the recording paper will already have it The grid marking setup can be done using output format select...

Страница 101: ...1 channel of rhythm will be recorded for 10 seconds at the bottom 6CH 1 10 second ECG will be recorded as I II III aVR aVL and aVF for the first 5 seconds and V1 V2 V3 V4 V5 and V6 for the next 5 sec...

Страница 102: ...output speed The set value will be displayed on the screen and when you choose 3CH 1 3 Channels 1 Rhythm Report will be shown at the top of the printing form 6CH 1 for 6 Channels 1 Rhythm Report 12CH...

Страница 103: ...V 1 31 2021 05 103 171 Doc BN OP C7 Print Line You can set the line thickness of the printouts Three setting options are available Thin Normal and Thick...

Страница 104: ...ll POP UP menu will appear at the bottom and it will disappear when choosing OFF Text Vector Guide ST Map Dot or ST Map Bar Default is set as OFF Beat report will be attached to the printing form of c...

Страница 105: ...can use all of the 3 leads while when 3CH 1 6CH 1 or Long Time is chosen you can use the first value only Default is set as II V1 V5 so when 3CH 1 6CH 1 or Long Time is chosen you can use the lead II...

Страница 106: ...show the outcome to the user This function can be turned On or Off in this window When the preview setting is set to On the Preview screen will appear following data collection with RECORD AUTO and NE...

Страница 107: ...conds signal data ID Name and Diagnosis The next page ECG 6 10 seconds signal data ID Name Filter setup Rhythm parameter values Note The preview screen will be shown with output speed set to 25mm sec...

Страница 108: ...setting up as ON diagnosis button should be used only after signalling is stabilized Upon ON setup diagnosis and output can be affected if the saved data is not stabilized this is because the data is...

Страница 109: ...ould help to reduce or remove those unwanted signals from the output of the device To change filter information enter the Filter Setup Menu either by moving the selection block over and select it usin...

Страница 110: ...used by the patient s breathing It can be seen as ECG signals being overlapped on a large parabola The Baseline filter provides 0 05Hz 0 1Hz and 0 2Hz and its set values will appear as Base 0 05Hz Bas...

Страница 111: ...z power noise correspondingly In the U S the AC filter should be set to 60Hz Check the properties of your power source before setting AC filter properties Under battery power there would be no AC nois...

Страница 112: ...50Hz and OFF option is offered Here 40Hz option means the filter will remove any unwanted signal with frequencies above 40Hz Low Pass Filter values will appear on the bottom of the printout form if LP...

Страница 113: ...ng the filter NOTE Applying filters may distort the original signal Therefore the proper use when the recording signal is in bad condition can reduce ECG signal distortion Referationlly it is recommen...

Страница 114: ...he patient and Leads is improper lead fault can occur In that case you can choose whether to have the lead fault message on the ECG Main page or not Selecting Lead Fault menu will switch between ON an...

Страница 115: ...ated If RA Lead becomes Fault Lead Fault message might not be shown and the wave types of all Leads might be indicated If LL Lead becomes Fault the wave type of II and V1 V6 Leads will not be indicate...

Страница 116: ...onitoring or recording You can select ON or OFF by clicking the Pacemaker item When it is on the location of the pacemaker will be printed out while when it is off it will not be printed out Note It i...

Страница 117: ...device to ring an alarm sound when a QRS beat has generated while it is in waiting mode Users can either set the alarm ON or OFF by selecting the menu with ON turning on and OFF turning off the alarm...

Страница 118: ...to connect electrodes on the graphic display Focusing on and selecting Leads Place menu will bring the screen shown below At that time enter the screen touch rotary or ESC key and then the how to con...

Страница 119: ...mode When it is set to ON 60bpm Sinus Normal Rhythm signal will be displayed on the device and DEMO sign will appear on the upper part of the LCD screen Users can test rhythm diagnostic copy printing...

Страница 120: ...consists of the AUTO key on the control panel setup item and auto diagnosis setup item AUTO key setting When the Store menu has the focus click it to set up Yes or No When Yes is selected you can ope...

Страница 121: ...he focus click it to set up Yes or No When Yes is selected you can operate the ECG test by pressing the AUTO key on the control panel and the ECG data will be automatically print out but when No is se...

Страница 122: ...transferred if is USB ECG data will be automatically saved to the external USB memory In addition if it is set up as Server data will be automatically transferred to PC Server IP among Network options...

Страница 123: ...d it will disappear when selecting one of OFF Standard or Professional When OFF is selected the content of diagnosis will not be printed out whereas when Standard is selected the basic information of...

Страница 124: ...nu window as below will appear and it will disappear when selecting one of AUTO 60msec or 80msec When AUTO is selected J Point associated with ST Level will be automatically set up at the time of prin...

Страница 125: ...f V3 V6 electrodes need to be changed from left to right Basically the V Change is set as None and when you click this menu a Leads Change POP UP window will show up as below Each time you select the...

Страница 126: ...lected as 1 3 5 or 10 min NOTE For Arrhythmia and HRV only 3 5 or 10 min are available Upon output of Arrhythmia and HRV ECG with pre setup duration will be printed out by one choice of Lead a line at...

Страница 127: ...7 171 Doc BN OP C7 Arrhythmia setting When performing Long Time recording arrhythmia printout options can be configured If Arrhythmia option is selected as ON printout will be performed if OFF it will...

Страница 128: ...2021 05 128 171 Doc BN OP C7 HRV setting When performing Long Time recording HRV printout options can be configured If HRV option is selected as ON printout will be performed if OFF it will not be per...

Страница 129: ...ctrodes or the patient cable WARNING The ECG patient cable provided is supposed to use for ECG measuring Do not use it to measure breathing Caution Use the electrodes same as the provided ones or with...

Страница 130: ...s selected the other Setup screen related to User Security Set and Manufacture Set will appear At User Security Set setup it is possible to change Factory Setting and User Password At Manufacture Set...

Страница 131: ...es are as follows Contents Setting Value Gain 10mm mV Speed 25mm s Start Option Main Unit Height cm Weight kg Date Type YYYY MM DD Patient Info File Delete Device Name Cardio7 Net Device Wired Configu...

Страница 132: ...Low Pass Filter 150Hz EMG Filter Off ECG Store Yes Print Out Yes Export None Diagnosis Professional ST Level Auto V Channel None Long Time Off Arrhythmia Off HRV Off Target PC Method Manual Format EKG...

Страница 133: ...V 1 31 2021 05 133 171 Doc BN OP C7 Data Erase All data in Files and Worklist will be deleted For the data erasure function a password must be entered...

Страница 134: ...7 Change PW User Password can be configured User Password is the number entered for Factory and Data Erase The password should be a 4 digit number NOTE The default User Password is 1234 If you forget...

Страница 135: ...21 05 135 171 Doc BN OP C7 Chapter 3 ECG file management 1 Display and function 2 Data printing 3 Data deleting 4 System Setup 5 Switching menu 6 Patient information 7 Data search 8 Data transfer 9 Da...

Страница 136: ...e among total 20 pages 1 Page contains 24 lines of Data If Rotary is selected the focus will be moved to Data list Button to skip to the previous page Button to skip to the next page Button to check t...

Страница 137: ...n be found on Print button When there is no patient information you want to select please press or button or the Search menu to move to other pages Menu direction Rotary Key s right spin Print Delete...

Страница 138: ...Select button one data will be deleted but if you press the All button all data in the list will be deleted In order to help the user not to make a mistake when deleting the data a message shown as b...

Страница 139: ...ECG When you click the menu Retain will be set up and when clicking the menu again Delete will be set up Default is set as Delete Delete To set up the patient information is deleted when you move to...

Страница 140: ...To save the data in the BMS Server or PACS Server USB To save the data in the external USB Memory When clicking the Method button a small line window will show up and it will disappear when selecting...

Страница 141: ...menu is used to select the user s options after the data has transferred Yes To delete the data after it has transferred No To not delete the data after it has transferred NOTE If Format menu has onl...

Страница 142: ...cting a saved data at the ECG file management page by using a rotary key or a touch screen the focus will appear at the data list Press the Info button Then you will see a Patient Information window a...

Страница 143: ...a Patient Search window as shown below If you rotate the rotary key the search condition button will have focus Press the Find button after inputting the information about ID Name Date or Age When the...

Страница 144: ...pressing the NETWORK key a message will show up asking which menu you want to select If you transfer a selected data a message with progress speed or other information will be shown as below If you tr...

Страница 145: ...data to PC immediately after measuring 2 When transmitting multiple sets of measurements rather than sending all data at once we recommend users to transmit individual records after each measurement N...

Страница 146: ...P C7 9 Data import You can import the saved data from the external device at the ECG data file management page by using the RECORD key on the control panel When pressing the RECORD key Import File Lis...

Страница 147: ...rompting you to select one of them will appear If the external device is not connected to Cardio7e Cardio7 the error message will be shown as below If you select the Selected menu in the Import File L...

Страница 148: ...Doc BN OP C7 If you select the All menu in the Import File List screen a system message with progress speed will be shown as below After receiving all data from the external device a system message w...

Страница 149: ...V 1 31 2021 05 149 171 Doc BN OP C7 Chapter 4 ECG worklist management 1 Display and function 2 ECG Test 3 Data deleting 4 System Setup 5 Switching menu 6 Patient information 7 Data search...

Страница 150: ...t page among total 20 pages 1 Page contains 24 Data If Rotary is selected the focus will be moved to Data list Button to skip to the previous page Button to skip to the next page Button to check the p...

Страница 151: ...earch menu to move to other pages Menu direction Rotary Key s right spin Exam Delete Setup Main Page information Info Search Exam Menu direction Rotary Key s left spin Exam Search Info Page informatio...

Страница 152: ...Select button one data will be deleted but if you press the All button every data in the list will be deleted In order to help the user not to make a mistake when deleting the data a message shown as...

Страница 153: ...Setup When clicking the Setup button at the ECG data worklist management page you will see the worklist System Setup window as shown below Note The steps for Network Hospital and Service Setting are i...

Страница 154: ...when moving to another main screen ECG skip to ECG main screen FILE skip to File Management screen Spiro skip to spirometry measurement screen 6 Patient information When selecting saved data at the E...

Страница 155: ...a Patient Search window as shown below If you rotate the rotary key the search condition button will have focus Press the Find button after inputting the information about ID Name Date or Age When the...

Страница 156: ...V 1 31 2021 05 156 171 Doc BN OP C7 Chapter 5 System Management 1 Maintenance and Cleaning 2 Regular Check up 3 Trouble Shooting 4 Manufacturer Declaration...

Страница 157: ...auze 40 C 104 F after use and clean it with clinical alcohol once a week Do not submerge the product or the ECG cable into any liquid or detergent Also make sure there is no liquid in the product or t...

Страница 158: ...out with noise First check that you set up the AC power filter If it still makes noise with the AC power supply please connect grounding line to the grounding electrode of the system The grounding sh...

Страница 159: ...t information window to implement 5 Wrong password Please re enter your password When wrong password is entered Check and input right password 6 Enter password length error Please re enter your passwo...

Страница 160: ...memory fails Check server PC s networking conditions and USB memory and retry 17 Check Worklist Server IP When server IP of PACS worklist is not input Input server IP of PACS worklist in turn of Syst...

Страница 161: ...se of printer for at least one minute and try again 29 Printer Paper empty When the printing paper is completely out Please put new roll of printing paper for use 30 Printer Printer not ready When the...

Страница 162: ...d Flicker Emission IEC 61000 3 3 Pst 1 Plt 0 65 Tmax 0 5 dmax 4 dc 3 3 Electrostatic Discharge Immunity IEC 61000 4 2 8 kV Contact 2 4 8 15 kV Air Radiated RF Electromagnetic Field Immunity IEC 61000...

Страница 163: ...ion electromagnetic emissions The Cardio7e Cardio7 is intended for use in the electromagnetic environment specified below The customer or the user of the Cardio7e Cardio7 should assure that it is used...

Страница 164: ...specified below The customer or the user of the Cardio7e Cardio7 should assure that it is used in such an environment IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment gu...

Страница 165: ...in UT for 5 s Mains power quality should be that of a typical commercial or hospital environment If the user of the Cardio7e Cardio7 requires continued operation during power mains interruptions it i...

Страница 166: ...tion applicable to the frequency of the transmitter Recommended separation distance Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m 80 MHz to 800 MHz 800 MHz to 2 5 GHz where P is the maximum...

Страница 167: ...radiated RF disturbances are controlled The customer or the user of the Cardio7e Cardio7 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF...

Страница 168: ...face Touch screen Rotary push knob Printer Resolution Thermal print head Roll Paper Report paper width A4 Letter 215mm 8 5 length A4 297 mm 11 7 Letter 279mm 11 Resolution 8dot mm 0 125mm pitch Patien...

Страница 169: ...scanner Safety Conformity Class I Type CF defibrillation proof applied part Environ mental Operation Ambient Temperature 10 40 Relative Humidity 30 85 RH Atmospheric Pressure 70 106KPa Storge Ship Am...

Страница 170: ...of purchase Date of purchase yyyy mm dd Customer Hospital Address Name Contact No Seller s Name Manufacturer s Name Thanks for purchasing Cardio7e Cardio7 This product is a medical machine This produ...

Страница 171: ...4 6638 Fax 1 714 734 1761 E mail support bionetus com Website www bionetus com European sales service representative Bionet Europe GmbH 2Li Bessemerstr 51 D 12103 Berlin Germany Tel 49 30 240 374 52 E...

Отзывы: