Select 'ON' when applying ‘EMG’ filter; otherwise, select 'OFF.' Setup values will be shown
on the bottom of the
printout form as ‘EMG:
On’ or ‘EMG:
Off’ when applying the filter.
NOTE
Applying filters may distort the original signal. Therefore, the proper use when the
recording signal is in bad condition can reduce ECG signal distortion.
Referationlly, it is recommended to apply the Baseline Drift filter and AC filter all times and
that the Near Field filter is used properly. It is also recommended that the LPF be set to
150 Hz for diagnostic outputs.