Doc. # : BN-OP-C7
2) Regular Check-up
Like every other medical product, Cardio7e/Cardio7 requires a regular check-up once
a year. Please refer to the Service manual for the information on the check-up.
3) Trouble Shooting
⚫
Printing was not successful:
The cover may not be closed properly. Please close the cover and print again.
⚫
When using the battery power, there was a beep for more than 3 times and a
message came as shown below:
The battery is low. Please connect to the AC power to use.
⚫
The signal is printed out with noise:
First, check that you set up the AC power filter. If it still makes noise with the AC
power supply, please connect grounding line to the grounding electrode of
the system.
The grounding should not be used with the AC power’s grounding and please
connect it to
the metal around the patient’s bed or the building.
⚫
When The external USB keyboard is not working properly:
Connect the external USB keyboard to the USB port before you turn on the
power of Cardio7e/Cardio7. If you connect the external USB keyboard to the
USB port after you turn on the power of Cardio7e/Cardio7, you should turn off
the power of Cardio7e/Cardio7 and turn on again.
If it is still not working, please contact our customer service center.
Содержание Cardio7
Страница 21: ...V 1 31 2021 05 21 171 Doc BN OP C7 Front View Printer Cover Switch...
Страница 61: ...V 1 31 2021 05 61 171 Doc BN OP C7 Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 62: ...V 1 31 2021 05 62 171 Doc BN OP C7 Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 63: ...V 1 31 2021 05 63 171 Doc BN OP C7 Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 64: ...V 1 31 2021 05 64 171 Doc BN OP C7 Record Report Form 12 CH Record Report Form 6 CH 1 RHY ST Map...
Страница 65: ...V 1 31 2021 05 65 171 Doc BN OP C7 Record Report Form BEAT REPORT TEXT Record Report Form BEAT REPORT VECTOR...
Страница 67: ...V 1 31 2021 05 67 171 Doc BN OP C7 Record Report Form BEAT REPORT ST Map Bar...