V 1.31
2021. 05
105
/
171
Doc. # : BN-OP-C7
[Rhythm]
You can set up the rhythm leads for printing out the diagnosis of 3CH+1, 6CH+1, 3CH+3
and Long Time (1,3,5,10
Minute). You can choose 3 rhythm leads and when ‘3CH+3’ is
chosen, you can use all of the 3 leads, while when ‘3CH+1’, ‘6CH+1’ or ‘Long Time’ is
chosen, you can use the first value only.
Default is set as ‘II, V1, V5’, so when 3CH+1, 6CH+1 or Long Time is chosen, you can use
the
lead ‘II’ and 3CH+3 for lead ‘II, V1, V5’.
If you choose ‘Rhythm’ menu, ‘Rhythm Channel’ window will be displayed as shown
below.
At that time,
select either the ‘CH1,’ ‘CH2’ or ‘CH3’ menu; and then choose one from
Lead I ~ V6 in the select menu.