V 1.31
2021. 05
43
/
171
Doc. # : BN-OP-C7
Countermeasures for Bad Lead Connection
After turning on the power switch, enter the ECG test to check waveforms of all leads.
On this menu, operators can check waveforms of all leads and their noise and may see
“Lead Fault” message displayed on the LCD screen.
NOTE
-
Message will be displayed when Lead fault is set to ON in System Setup and
will NOT be displayed when set to OFF.
-
Please conduct an ECG test with the lead connection is correct.
The following two cases could cause a lead fault error.
- First case,
leads are detached from the patient’s body. In this case, re
-attach leads
normally.
-
Second case, signal conductivity between leads and the patient’s body is low. In this
case, replace the electrodes.
If the two above cases have been performed and the operator is still experiencing noise
or lead fault conditions, the patient cable may be faulty. Please contact the Bionet
service center.
Lead Fault! All Leads.
Содержание Cardio7
Страница 21: ...V 1 31 2021 05 21 171 Doc BN OP C7 Front View Printer Cover Switch...
Страница 61: ...V 1 31 2021 05 61 171 Doc BN OP C7 Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 62: ...V 1 31 2021 05 62 171 Doc BN OP C7 Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 63: ...V 1 31 2021 05 63 171 Doc BN OP C7 Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 64: ...V 1 31 2021 05 64 171 Doc BN OP C7 Record Report Form 12 CH Record Report Form 6 CH 1 RHY ST Map...
Страница 65: ...V 1 31 2021 05 65 171 Doc BN OP C7 Record Report Form BEAT REPORT TEXT Record Report Form BEAT REPORT VECTOR...
Страница 67: ...V 1 31 2021 05 67 171 Doc BN OP C7 Record Report Form BEAT REPORT ST Map Bar...