Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-4-
11
<5%UT
(>95%dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5, 6 cycles
70 %UT
(30 %dip in UT)
for 25,30 cycles
<5%UT
(>95%dip in UT)
for 5 s
<5%UT
(>95%dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5, 6 cycles
70 %UT
(30 %dip in UT)
for 25,30 cycles
<5%UT
(>95%dip in UT)
for 5 s
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the
Cardio7e/Cardio7 requires
continued operation during
power mains interruptions,
it is recommended that the
Cardio7e/Cardio7 be
powered from an
uninterruptible power supply
or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE
U
T
is the a. c. mains voltage prior to application of the test level.
Содержание Cardio7
Страница 21: ...V 1 31 2021 05 21 171 Doc BN OP C7 Front View Printer Cover Switch...
Страница 61: ...V 1 31 2021 05 61 171 Doc BN OP C7 Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 62: ...V 1 31 2021 05 62 171 Doc BN OP C7 Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 63: ...V 1 31 2021 05 63 171 Doc BN OP C7 Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 64: ...V 1 31 2021 05 64 171 Doc BN OP C7 Record Report Form 12 CH Record Report Form 6 CH 1 RHY ST Map...
Страница 65: ...V 1 31 2021 05 65 171 Doc BN OP C7 Record Report Form BEAT REPORT TEXT Record Report Form BEAT REPORT VECTOR...
Страница 67: ...V 1 31 2021 05 67 171 Doc BN OP C7 Record Report Form BEAT REPORT ST Map Bar...