V 1.31
2021. 05
111
/
171
Doc. # : BN-OP-C7
There are three types of an AC power noise filter: 60Hz, 50Hz and OFF. When set to OFF,
the filter will not be applied. If the filter is set to 50Hz or 60Hz, then the AC filter will
remove 50Hz or 60Hz power noise correspondingly. In the U.S., the AC filter should be
set to 60Hz. Check the properties of your power source before setting AC filter
properties.
Under battery power, there would be no AC noise being generated. In this case,
please set the AC filter setting to OFF so that output signal would not be distorted.
The status of the AC filter can be checked on the bottom part of the printout -
indicating AC 60Hz, AC 50Hz or AC OFF
Содержание Cardio7
Страница 21: ...V 1 31 2021 05 21 171 Doc BN OP C7 Front View Printer Cover Switch...
Страница 61: ...V 1 31 2021 05 61 171 Doc BN OP C7 Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 62: ...V 1 31 2021 05 62 171 Doc BN OP C7 Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 63: ...V 1 31 2021 05 63 171 Doc BN OP C7 Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 64: ...V 1 31 2021 05 64 171 Doc BN OP C7 Record Report Form 12 CH Record Report Form 6 CH 1 RHY ST Map...
Страница 65: ...V 1 31 2021 05 65 171 Doc BN OP C7 Record Report Form BEAT REPORT TEXT Record Report Form BEAT REPORT VECTOR...
Страница 67: ...V 1 31 2021 05 67 171 Doc BN OP C7 Record Report Form BEAT REPORT ST Map Bar...