7
Installation
To remove the Tip, repeat this process in reverse order.
1. Slide the Handpiece laterally toward the Tip Remover or revolving Tip
Holder.
2. Place your thumb against the selected Tip slot to prevent laser Tip from
falling out of the Tip Holder when disconnecting it from the Handpiece.
3. Carefully lift the Handpiece to disengage the Tip ferrule from the
Handpiece head.
4. Use tweezers to slide the Tip out from the Tip Holder or Tip Remover;
dispose of the used Tip in a medical waste sharps container.
NOTE:
If the laser cuts hard- and soft-tissue slower than expected
after installing the Fiber Cable, please follow the table in Section 11,
Troubleshooting the Delivery System.
CAUTION:
If using the Turbo Handpiece (p/n
6201126
), use the
same techniques when installing or removing Turbo Tips; however,
Turbo Tips do require a different Tip Holder. Not that the Turbo Tip
Holder/Remover tool (p/n 7200407) ONLY works with Turbo Tips;
the tool used for the Gold Handpiece DOES NOT work with Turbo
Tips. Refer to the Turbo Handpiece instructions for use for more
information.
Fig 7.28
Fig 7.29
Fig 7.30
Fig 7.31
Fig 7.32
30
Waterlase iPlus User Manual
NOTE:
Turbo Tips are unique to the Turbo Handpiece and are not
interchangeable with other tips and not compatible with any
Handpiece other than the Turbo Handpiece.
NOTE:
The Tip must be sterilized before initial use and between
patients. If not sterilized in the Tip Holder, remove the Tip from its
sterilization pouch and insert it into the Tip Remover or the
revolving Tip Holder by aligning the first groove of the Tp ferrule
against the receiving edges of the holder, then sliding the Tip in;
using tweezers facilitates this process (Fig. 7.28).
3. Align the Tip orifice of the Handpiece over the input end of the Tip
placed in the Tip Remover or revolving Tip Holder (Fig. 7.29).
4. Carefully lower the Handpiece and insert a clean/inspected tip all the way
until the shoulder of the Tip ferrule sits against the Handpiece head
(Fig. 7.30, 7.31).
5. Slide the Handpiece laterally away from the Tip Remover or Tip Holder
(Fig. 7.32).
Содержание Waterlase iPlus
Страница 1: ...User Manual...
Страница 2: ...PAGE INTENTIONALLY LEFT BLANK...
Страница 62: ...Appendix A Labels Laser Explanatory Label Laser Aperture Label 62 Waterlase iPlus User Manual...
Страница 78: ...Appendix G Electromagnetic Compatibility 78 Waterlase iPlus User Manual...
Страница 79: ...79 Appendix G Electromagnetic Compatibility Waterlase iPlus User Manual...
Страница 80: ...Appendix G Electromagnetic Compatibility 80 Waterlase iPlus User Manual...
Страница 81: ...PAGE INTENTIONALLY LEFT BLANK...