Appendix D
Tip Inspection
6. Remove the Tip from the test holder using the Tip
Remover. If the Tip is contaminated at either end, try
cleaning it as shown below. If the Tip is damaged,
replace it from the Handpiece using the Tip Remover
and dispose of it.
To replace the batteries for the built-in microscope light, gently pull apart the upper and lower tubes of
the microscope. Locate the battery cover marked with “OPEN”, slide the cover in the direction of the
arrow, remove the old batteries and replace with two size AA 1.5 volt (Europe size M) batteries.
Tip Remover with Tip
inside
1. Remove the Tip from the Handpiece and insert it into
the correct side of the Tip test holder as shown using
the Tip Remover.
2. Insert the Tip test holder into the test adapter with
the distal (or laser-emitting) end of the Tip toward the
microscope.
3. Slide the adapter over the microscope to move the Tip
surface toward the focal point of the microscope. The
focal point lies in the plane at the end of the clear end
tube of the microscope.
4. Turn on the microscope’s built-in light by gently pulling
apart the upper and lower tubes, or hold it up to
another light source, and bring the surface of the Tip
into focus using the thumbwheel. Examine the Tip
surface carefully for damage or contamination.
5. To examine the proximal (or Fiber Cable) end of the Tip,
remove the adapter from the microscope, and gentl��
t
the other side of the test holder into the clear end tube
of the microscope. Refocus the Microscope.
Tip Remover with Tip inside
TIP INSPECTION INSTRUCTIONS
Focal Point
Thumbwheel
Good
Burnt
Broken
Contaminated
66
Waterlase iPlus User Manual
Содержание Waterlase iPlus
Страница 1: ...User Manual...
Страница 2: ...PAGE INTENTIONALLY LEFT BLANK...
Страница 62: ...Appendix A Labels Laser Explanatory Label Laser Aperture Label 62 Waterlase iPlus User Manual...
Страница 78: ...Appendix G Electromagnetic Compatibility 78 Waterlase iPlus User Manual...
Страница 79: ...79 Appendix G Electromagnetic Compatibility Waterlase iPlus User Manual...
Страница 80: ...Appendix G Electromagnetic Compatibility 80 Waterlase iPlus User Manual...
Страница 81: ...PAGE INTENTIONALLY LEFT BLANK...