8
Operating Instructions
Changing the Water in the Bottle
When the system detects that the water level in the self-contained bottle is low, a blinking button
with a low water level symbol will appear next to the Settings button. Place the system into
Standby
mode and follow the steps outlined in Section 7: Filling the Self-Contained Water System Bottle. To
return to
Ready
mode, press the Main Function button below the touchscreen.
NOTE:
When the bottle is disconnected, an Error screen will
appear. When the bottle is re-attached, the Error screen will
clear automatically when the system checks the bottle status
(approximately 5 seconds).
Three setting categories are shown at the bottom of the screen, i.e., Handpiece and Tip, Laser, and
Spray. From here, adjustments can be made to the following parameters:
•
Handpiece
type and
Tip
type (only the Tips which are allowed for this procedure)
•
Laser Power
, Pulse Repetition Rate, and Pulse Mode
•
Water Spray
and Air Percentage setting
Handpiece
The Handpiece recommended for the current procedure step appears at the bottom of the screen.
However, the Handpiece can be changed by taking the following steps:
1. Press the image of the Handpiece that the user wants to install; a progress timer will appear in
the form of a segmented circle outside of the Handpiece icon and the system will automatically
purge the water from the installed Handpiece (Patient Air 100% ON, air pressure in the bottle
OFF, Patient Water 100% ON). This step will take 3 – 4 seconds.
2. When the water is purged from the Handpiece, a new message will appear on the screen:
“Exchange Handpiece now and then select Tip.” Attach the new Handpiece to the Fiber Delivery
System.
3. Press the image of the Tip chosen for the procedure; patient air (and internal cooling air) will be
activated through the Handpiece, and a new Tip may now be inserted.
4. Once a new Tip is attached to the Handpiece, press the Handpiece image again; a progress
timer will again re-appear for 3-4 seconds, and the Handpiece will be primed with water.
To return to the Procedure screen, press either the Back or Handpiece button located at the
bottom of the display.
34
Waterlase iPlus User Manual
Содержание Waterlase iPlus
Страница 1: ...User Manual...
Страница 2: ...PAGE INTENTIONALLY LEFT BLANK...
Страница 62: ...Appendix A Labels Laser Explanatory Label Laser Aperture Label 62 Waterlase iPlus User Manual...
Страница 78: ...Appendix G Electromagnetic Compatibility 78 Waterlase iPlus User Manual...
Страница 79: ...79 Appendix G Electromagnetic Compatibility Waterlase iPlus User Manual...
Страница 80: ...Appendix G Electromagnetic Compatibility 80 Waterlase iPlus User Manual...
Страница 81: ...PAGE INTENTIONALLY LEFT BLANK...