10
3
Contraindications, Warnings, and Precautions
CONTRAINDICATIONS
All clinical procedures performed with the
Waterlase
iPlus must be subjected to the same clinical judgment
and care as with traditional techniques. Patient risk must always be considered and fully under stood
before clinical treatment. The clinician must completely understand the patient’s medical history prior to
treatment. Exercise caution for general medical conditions which might contraindicate a local procedure.
Such conditions may include, but are not limited to, allergy to local or topical anesthetics, heart disease
(e.g. pacemakers, implantable defribulators), lung disease, bleeding disorders, or an immune system
defi
-
ciency. Medical clearance from the patient’s physician is advisable when doubt exists regarding treatment.
WARNINGS AND PRECAUTIONS
PRESCRIPTION STATEMENT
Federal Law restricts this device to sale by or on the order of a
dentist or other
licensed
dental
practitioner.
EYEWEAR
D
octor, patient, assistant, and all others insi
d
e or entering the operatory must always wear
appropriate laser protection eyewear for the 2780nm
wavelength, OD
4 (DI LB4)
or greater.
Always
check the eyewear specifications imprinted on the frame of the glasses to
ensure they offer the
required protection for the specific laser wavelength.
Prior to use, inspect eyewear for pitting a cracking.
Replace if damaged; do not use.
TRAINING
Only licensed professionals who have reviewed and understood this user manual, and have been
trained on how to correctly operate the system should use this device. Surgical procedures related to
soft-tissue, osseous, endodontic, or periodontal surgery should only be performed by clinicians who
have training and experience in Oral Maxillofacial, Periodontal, or Endodontic Surgery.
ANESTHESIA
Although in most cases anesthesia may not be required, patients should be closely monitored for
signs of pain or discomfort. If such signs are present, adjust settings, apply anesthesia or cease
treatment, if required.
TREATMENT, TECHNIQUE AND SETTINGS
Only licensed professionals who have reviewed and understood this user manual should use this
device.
Always start treatment at the lowest power setting for the specific tissue and increase as
required.
Observe clinical effects and use judgment to determine the aspects of the treatment
(technique, proper power, pulse mode, air and water settings, tip type and duration of operation) and
make appropriate power, air and water adjustments to compensate for varying tissue composition,
density and thickness.
Waterlase
iPlus User Manual
Содержание Waterlase iPlus
Страница 1: ...User Manual...
Страница 2: ...PAGE INTENTIONALLY LEFT BLANK...
Страница 62: ...Appendix A Labels Laser Explanatory Label Laser Aperture Label 62 Waterlase iPlus User Manual...
Страница 78: ...Appendix G Electromagnetic Compatibility 78 Waterlase iPlus User Manual...
Страница 79: ...79 Appendix G Electromagnetic Compatibility Waterlase iPlus User Manual...
Страница 80: ...Appendix G Electromagnetic Compatibility 80 Waterlase iPlus User Manual...
Страница 81: ...PAGE INTENTIONALLY LEFT BLANK...