Chapter 6 Parameter Monitoring
User Manual of Veterinary Monitor
--77--
6.7.2 Safety Information
Warning
Pressure sensors specified in this User manual only are allowed to be used and repeated use of
disposable pressure sensors is not allowed.
Contact of the monitor
‟s sensors and cables with high-frequency surgical equipment should be
avoided in order to prevent animals from being burnt by leaked current when the monitor is in
contact with high-frequency surgical equipment.
Contact of accessories with electrically connected metallic parts should be avoided while
connecting and using such accessories.
The working temperature of such accessories should be considered when they are used; and
please refer to the Accessory Use Manual for more details.
If the packaging of the invasive pressure sensor marked for disinfection is damaged, please stop
using it and contact the provider of the invasive pressure sensor.
If the pressure sensor and conduit are damped, please stop using it and contact the provider of the
invasive pressure sensor.
Attention
Before monitoring, please zero-calibrate first.
Make sure that the pressure sensors are constantly leveled with the heart during monitoring.
Please continuously inject heparin saline to flush conduits to prevent them from being clogged and
maintain the pressure-measuring paths are smooth while connecting the conduits firmly to prevent
them from any displacement or separation that would affect the IBP measure.
6.7.3 Monitoring Steps
1
Insert the pressure transducer extension cables into the IBP interface on the monitor‘s left panel.
2
Prepare flushing solution.
3
Flush the system and deflate the conduits completely. Make sure no bubbles exists in the sensors and valves.
4
Connect the pressure tube and animal conduit.
5
Place the sensor on the same level with the heart, approximately at the midaxillary line.
6
Select correct label.
7
Zero-calibrate the sensors. Close the sensor‘s valves toward the air and open the valve in connection with the
animal when the zero-calibration is completed.
Figure 6.37 IBP Monitor Connection Diagram
Содержание iM 12
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Страница 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Страница 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Страница 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Страница 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Страница 194: ...Appendix E EMC Guidance and Manufacture s Declaration 182 User Manual of Veterinary Monitor The Blank Page ...
Страница 195: ......