Appendix B Product Specifications
User Manual of Veterinary Monitor
--149--
Rated voltage
14.8 V
Battery Capacity
4.4 Ah
Length of Power Supply
In environment temperature ranging from 20
℃
to 30
℃
and in standard configuration
(the SpO
2
sensor connects, the ECG cable and Temp cable disconnect, the ―Measure
Mode‖ of NIBP is ―Auto‖ and the ―Interval‖ is 15 minutes), the continuous working
time of a single battery is not less than 5 hours.
Time for recharging battery to 90%
from zero power state
In environment temperature ranging from 20
℃
to 30
℃
and with the machine
turning off, the charging time is not more than 12 hours to charge the battery to 90%.
Shutdown Delay
0 s, 0.5 s, 1 s, 1.5 s, 2 s
Host LED
Physiological Alarm Indicator Lamp
1 (Dual color yellow & red)
Technical Alarm Indicator Lamp
1 (Blue)
Power Switch Indicator Lamp
1 (Green)
AC Power Indicator Lamp
1 (Green)
Battery Power Indicator Lamp
1 (Green)
Battery Charging Indicator Lamp
1 (Green) (Only for 12 inches monitor)
Keypad Backlight
5 (White)
Alarm Pause Key Backlight
1 (Red)
Speaker
Give out alarm sound (45 dB
~
85 dB), keystroke sound and QRS sound.
Alarm sound complies with IEC 60601-1-8
Interface
Power
1 AC power port
Network
Standard RJ45 network port, which can network with the central monitoring system and
transmit all the Veterinary Monitored data to the central monitoring system.
USB
USB disk supported. For the manufacturer to upgrade and service the application
software, and export data (Structurally 2 USB host interfaces supported)
VGA
Supported, for connection of external display
Analog Output Port
1 piece. It can be connected to oscilloscope for output of the analog signals.
Nurse
Call
System
Interface
and
Defibrillation Synchronization Interface
1 piece. It can be connected to port of the nurse call system or the defibrillator.
Equipotential Terminal Port
1 piece
ECG Analog Signal Output
Bandwidth (-3 dB, reference 10Hz)
Surgery mode: 1 Hz
~
15 Hz
Monitor mode: 0.5 Hz
~
40 Hz
Diagnose mode: 0.05 Hz
~
150 Hz
Max. Transmission Delay
25ms (Wave filter closed under diagnose mode)
Sensitivity
1 V/mV ±5%
Accuracy of input signal reproduction
Using the method described in 4.2.7.1 of AAMI EC11 to test the overall system error,
which is within
±
5%;
Using method A and D described in 4.2.7.1 of AAMI EC11 to test frequency response.
Because of sampling characteristics and the asynchronism between sample rate and
signal rate of the ECG module, digital systems may produce a noticeable modulating
effect from one cycle to the next. This phenomenon, which is not physiologic, shall be
clearly described in the operator‘s and service manuals.
IBP Analog Signal Output
Bandwidth (-3 dB, reference 10Hz)
0 Hz
~
50 Hz
Max. Transmission Delay
30 ms (Filter closed)
Sensitivity
0.01 V/mmHg±5%
Содержание iM 12
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Страница 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Страница 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Страница 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Страница 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Страница 194: ...Appendix E EMC Guidance and Manufacture s Declaration 182 User Manual of Veterinary Monitor The Blank Page ...
Страница 195: ......