Chapter 6 Parameter Monitoring
--92--
User Manual of Veterinary Monitor
[Stop]: Select [Stop] to stop the current measurement if the measurement is too long.
[Cancel]: During the current measurement, select [Cancel] to cancel the measurement output and stop
the current measurement. After a measurement, select [Cancel] to delete the measured output.
[Record]: Select [Record] to print out the most recently-measured output.
[X-Scale]: Select [X-Scale] to adjust the scale of the X-axis. Options are 30 s and 60 s.
[Y-Scale]: Select [Y-Scale] to adjust the scale of the Y-axis. Options are 0.5º
C, 1º
C and 2.0º
C.
[Setup >>]: Select [Setup >>] to access the [C.O. Setup] menu.
E.
Averaged C.O. and C.I.
F.
Measurement windows
5.
When you see the message [Ready for New Measurement], select the [Start] button and then inject the
solution within 4 seconds, and the curve will be displayed. After the measurement, the curve and results will
enter measurement windows. At last, the monitor will prompt you to wait for a certain period of time before
starting a new measurement.
6.
Repeat step 5 until you decide to finish. Select by checking under the measurement windows, and the
monitor will automatically calculate and display the averaged C.O. and C.I..
If you perform more than six measurements, the oldest output will be deleted and the new output will enter.
Before an injection, turn off the stopcock to the PA catheter and turn on the stopcock to the injectable solution,
and then draw the injectable solution into the syringe.
When the injectable solution enters the syringe, turn off the stopcock to the injectable solution and turn on the
stopcock to the PA catheter, and then start an injection.
After the measurement is completed, turn off the stopcock to the PA catheter and turn on the stopcock to the
injectable solution.
Attention
During the C.O. measurement, blood temperature alarms are inactive.
6.9.4 Understanding the C.O. Display
The C.O. measurement is displayed on the monitor as numeric C.O. and TB in the C.O. parameter window as
shown below. To enter the [C.O. Setup] menu, select the C.O. parameter window.
6.9.5 Changing C.O. Settings
Setting the Temperature Unit
In the [User Maintain]
menu, select [Temp Unit] to toggle between [º
C] and [º
F].
Setting the C.O. Alarm
In [C.O. Setup] menu, select [Alarm Setup>>]
to enter the C.O. alarm window to set the alarm.
Temperature unit
Blood temperature
Cardiac output
Содержание iM 12
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Страница 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Страница 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Страница 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Страница 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Страница 194: ...Appendix E EMC Guidance and Manufacture s Declaration 182 User Manual of Veterinary Monitor The Blank Page ...
Страница 195: ......