Appendix C Alarm Information
User Manual of Veterinary Monitor
--163--
Appendix C Alarm Information
Physiological Alarm Information
Physiological Parameters
Alarm Information
Triggering Condition
Treatment Measure
xx Too High
xx value exceeds the alarm upper limit.
Check the physiological condition of the animal
and confirm if the setting of animal type and alarm
limit is suitable to the animal.
xx Too Low
xx exceeds the alarm lower limit.
Attention: xx represents the physiological parameter or name of a module, e.g. HR, ST-
Ⅰ
, SpO
2
, NIBP Systolic Blood Pressure and
RR, AG, etc.
ECG
Alarm Information
Triggering Condition
Treatment Measure
ECG Signal Too Weak
The animal ECG signal is too weak.
Check the animal state, electrode and lead cable.
Asystole
Heart beat NOT detected when preset cardiac
arrest threshold time has passed
If the animal suffers arrhythmia, check the animal
state, electrode and lead cable. Check if the setting
of arrhythmia trigger threshold is suitable to the
animal.
VFib/VTac
Fibrillating
waves
last
consistently
for
6s//Dominant rhythm of the adjacent ventricular
heart beats (V) and the heart rate is greater than
the upper limit of ventricular tachycardia
Extreme Tachycardia
Heart rate exceeds extreme tachycardia threshold
Extreme Bradycardia
Heart rate lower than extreme bradycardia
threshold
Ventricular Rhythm
Lead rhythm of the adjacent ventricular beat
exceeding the number of idioventricular rhythm
threshold, and the heart rate is lower than VT
rate.
Ventricular Bigeminy
Rhythm N, V, N and V
Ventricular Trigeminy
Rhythm N, N, V, N, N, V
Irregular Rhythm
Continuous irregular rhythm
PVCs/min
PVCs/min exceeds preset higher limit
Run PVCs
>
2
More than 2 continuous PVCs in the last minute
Couplet PVCs
Paired PVCs detected in the last minute
R on T
R on T detected within the last minute
Multiform PVCs
Ventricular premature of 2 or more forms is
detected in the last minute
HeartBeat Pause
Not detecting heart pacing within preset cardiac
arrest threshold time
Missed Beats
Unable to detect the heart pacing within 1.75
times of the mean RR period when the heart rate
is <100, or unable to detect the heart pacing in 1s
when the heart rate is >100.
Pacemaker NOT Capture
Asystole with pace-making pulse in the last
minute (Only applicable to pacemaker-wearing
animals)
The pacemaker has problem. Please check the
pacemaker.
Pacemaker Not Pace
No pace-making pulse detected within a period
that is 1.75 times the average R-R intervals (Only
applicable to pacemaker-wearing animals)
Содержание iM 12
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Страница 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Страница 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Страница 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Страница 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Страница 194: ...Appendix E EMC Guidance and Manufacture s Declaration 182 User Manual of Veterinary Monitor The Blank Page ...
Страница 195: ......