Appendix C Alarm Information
--166--
User Manual of Veterinary Monitor
CO
2
Sensor Preheating
The CO
2
sensor module is started and being
preheated.
Wait
CO
2
Zero Progress
CO
2
Being Zero Calibration
Wait
CO
2
Zero
Base
Inaccurate, Please Zero
The CO
2
reading is incorrect
Ensure if the airway is correctly connected. Carry out
zero calibration after the sensor temperature is
stabilized.
CO
2
Replace Adapter
CO
2
Requiring Oxygen Range Calibration
Please execute one calibration operation.
CO
2
Sensor
Software
Error
The CO
2
module has failed
Reinsert the module or restart the monitor.
CO
2
Airway Adapter Off
The airway adaptor is abnormal.
Check the airway and eliminate the blocking.
CO
2
Pump Shut
CO
2
Pump Closed
Confirm if CO
2
pump is closed.
CO
2
Calibrate Error
The CO
2
calibration is wrong
Recalibrate.
C.O.
Alarm Information
Triggering Condition
Treatment Measure
TB Sensor Off
The sensor is not connected or incorrectly
connected.
Check the sensor connection. Reconnect it.
IBP
Alarm Information
Triggering Condition
Treatment Measure
xx Sensor Off
(xx refers to an IBP label)
The sensor is not connected or incorrectly
connected.
Check the sensor connection. Reconnect it.
Others
Alarm Information
Triggering Condition
Treatment Measure
Recorder Initial Error
Recorder initialization error
Restart the equipment
Recorder Out of Paper
Recorder no paper or paper position wrong
Check the print paper and reinstall it.
Recorder Serial Error
The recorder serial port communication has an
error
Clear the print task and restart the equipment
Recorder Uninstall
The recorder is improperly installed
Check the recorder installation and restart the
equipment.
Head of Print Hot
The recorder has worked too long
Clear the print task and output the records after the
machine has cooled.
Voltage of Battery Too
Low
The battery voltage is low and cannot maintain
long-time monitoring
Switch to AC power supply. Power supply by battery
can only be used when the battery is fully recharged.
Very
Low
Voltage,
Shortly Logout
The battery voltage is too low. To avoid data loss
due to low power, the system will soon activate
the automatic shutdown procedure.
Switch to AC power supply. Power supply by battery
can only be used when the battery is fully recharged.
AG
Alarm Information
Triggering Condition
Treatment Measure
AG Sensor Off
AG sensor is not connected
Confirm that the AG sensor has been correctly
connected.
AG Sensor Too Hot
AG sensor temperature is too high
Check and stop using or change the sensor.
AG Sensor Too Cold
AG sensor temperature is too low
AG Pressure Too High
Airway pressure is abnormal
Check the animal and airway connection. Then restart
the monitor.
AG Pressure Too Low
AG
Airpressure
Too
High
The monitor environment affects the pressure
Check the airway connection and confirm if the
environment conforms to the monitor specifications
and if there is any special factor affecting the
environment pressure.
AG Airpressure Too Low
Содержание iM 12
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Страница 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Страница 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Страница 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Страница 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Страница 194: ...Appendix E EMC Guidance and Manufacture s Declaration 182 User Manual of Veterinary Monitor The Blank Page ...
Страница 195: ......