Chapter 6 Parameter Monitoring
--58--
User Manual of Veterinary Monitor
Resp Lead Setup
Select [Resp Lead]: I and II. It allows selecting the desired lead to meet different clinical demands for
optimal respiratory waves.
Attention
In order to obtain optimal respiratory waves, RA and LA electrodes should be kept level when I-lead
is selected for resp measure, and RA and LL electrodes should be diagonally kept when II-lead is
selected.
Detect. Mode Setup
Select [Detect. Mode]: Auto and Manual. Respiratory wave calculation mode can be set.
Auto Mode
In this mode, the monitor will automatically adjust the detection level according to waveform height and
whether cardiac artifact is existent, but the detection level dotted line is not displayed on the respiratory
waveforms.
[Auto] mode is recommended when respiratory rate is not close to the heart rate or animals are actively
respiration.
Manual Mode
In the Manual mode, users are required to set the respiratory detection level. Select [Upper Line] and
[Lower Line] to manually relocate the detection level dotted line on the respiratory waveforms. Once
determined, the detection level will not automatically adapt to different depths of respiration. Thus,
users are required to readjust the detection level to adapt to changes in the depths of respiration
according to actual conditions.
Users can directly click
and
in the window to raise or lower the upper/lower dotted lines if
the touch screen function is configured. When the upper/lower dotted lines of the respiratory waves are
blocked by [Resp Setup] Window, users can click the title panel on the window to drag it until the
upper/lower dotted lines can be relocated.
Attention
In case the respiratory waves are not displayed on the respiratory oxygenation diagram interface or
main screen, the detection level is still not adjustable even if the [Detect. Mode] is set Manual.
Respiration Alarm Parameter Setup
Select [Alarm Setup>>] to enter the menu [PAR.Alarm], then users can set RR alarm attributes, including:
Alarm Switch, Alarm Higher/Lower Limit, Alarm Level and Alarm Record On/Off.
6.3 PR
6.3.1 PR General Description
The mechanical movement of the heart causes pulsation of arteries, and the PR (pulse rate) can be obtained by
measuring this pulsation. The PR numerical value can be obtained through SpO
2
or any arterial pressure (IBP).
Содержание iM 12
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Страница 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Страница 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Страница 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Страница 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Страница 194: ...Appendix E EMC Guidance and Manufacture s Declaration 182 User Manual of Veterinary Monitor The Blank Page ...
Страница 195: ......