Chapter 6 Parameter Monitoring
User Manual of Veterinary Monitor
--49--
Other Setup
Select [ECG Setup]-[Others>>]:
QRS Volume:
0-10.0 and 10 indicate mute and the maximum volume, respectively.
Notch Filter:
when the Filter Mode is ‗Monitor‘ or ‗Surgery‘, the notch filter is defaulted open (50Hz or
60Hz); when the Filter Mode is ‗Diagnose‘, the notch filter can be opened or closed to meet different
demands.
Lead Set:
3-Lead, and 5-Lead.
Screen:
select ECG working interface, and different options are available for different [Lead Set] settings:
1)
3-Lead: Normal.
2)
5-Lead: Normal, 7 Lead Half and 7 Lead Full.
Paced:
please refer to Check on Pacemaking Status for more details.
Save Curve:
I, II, III, aVR, aVL, aVF and V, namely, the ECG lead which allows selection of Long ECG
review storage.
ST Use:
ST Point, J+60 and J+80, and ST use mode can also be set.
Smart Lead Off:
in case that Smart Lead Off is selected, when the lead that is set in ECG or being displayed
on the interface is disconnected but there are other leads still useable, the system will automatically select the
useable leads and re-compute the heart rate. Still, when it is reconnected, the now connected lead will be
automatically restored to its original state.
6.1.8 ST Analysis
The ST analysis function is not suitable for
<
10 kg.
ST analysis can measure ST elevation or reduction on a given lead.
Unit of ST measured value: mV or mm.
Meaning of ST measured value: a positive number indicates an elevation while a negative one indicates a
reduction.
ST Measuring Range: (-2.0 mV)
~
(+2.0 mV).
Warning
ST data should not be used as a judgment standard for doctors and its clinical significance is
decided by the doctors.
ST Analysis On/Off
Select [Main Menu]-[ECG Setup]-[ST Analysis>>] or select ECG waveform region or parameter region to
enter ST Analysis window. ST segment can be set [On] or [Off].
Filter Mode
ST Analysis function proceeds only when the Filter Mode is in the Diagnose Mode. ST Analysis function
will be automatically stopped when the Filter Mode is switched to ―Monitor‖ or ―Surgery‖ from ―Diagnose‖,
but the ST Analysis can still be started at this moment. Once it is started, the Filter Mode is automatically
switched to the Diagnose Mode; ST Analysis function is still stopped but can be manually started if the Filter
Mode is switched from Monitor or Surgery to Diagnose.
Содержание iM 12
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 22: ...Chapter 1 Operation Safety Information 10 User Manual of Veterinary Monitor The Blank Page ...
Страница 44: ...Chapter 3 Basic Operation 32 User Manual of Veterinary Monitor The Blank Page ...
Страница 54: ...Chapter 5 User Interface 42 User Manual of Veterinary Monitor The Blank Page ...
Страница 134: ...Chapter 8 Freeze and Review 122 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 10 Recording 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 154: ...Chapter 14 Maintenance 142 User Manual of Veterinary Monitor The Blank Page ...
Страница 180: ...Appendix C Alarm Information 168 User Manual of Veterinary Monitor The Blank Page ...
Страница 194: ...Appendix E EMC Guidance and Manufacture s Declaration 182 User Manual of Veterinary Monitor The Blank Page ...
Страница 195: ......