![BIO RAD PTC-0240 Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/bio-rad/ptc-0240/ptc-0240_operation-manual_2753175094.webp)
Running Programs
Tech Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com
7-7
The Status window will display Block, Sample, and Lid temperatures corre-
lating to the program running on the block chosen in the block selection
menu. These temperatures represent real-time readings and correspond to
the values represented graphically in the above window. The time remaining
in the current Step, and the Remaining time in the program are displayed
along with the current Cycle number.
Tip:
Recall that convenient and rapid toggling between the Status and
Graphs windows can be achieved by selecting the button in the lower right of
the Status and Graphs windows respectively.
To simultaneously view the status of all blocks including the Block Name,
Block Status, the name of the Program Running, the Time Remaining in the
program, the Time Elapsed, and the User name, select the <View> menu and
then <Status All> from the drop-down list.
If the Block Status indicates that the program was completed with errors,
select <Error Log> from the <View> menu to view error messages.
To view a run log including the date and time that a program was initiated by
a user, select <Cycler Log> from the <View> menu.