DNA Engine Tetrad 2 Thermal Cycler Operations Manual
10-12
Tech Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com
Table 10-2: Protocol Difficulty
Error Message
Cause and Result
Action
Reaction is working, but
broad low-molecular
weight band is seen
in gels.
“Primer-dimer” material often pro-
duces a broad band in the
<
100 bp
region of gels.
If obtaining appropriate reaction
product/s, no need to change
anything.
Minimize “primer-dimer” production
by designing primers with no 3’
self-complementary.
Reoptimize magnesium concentra-
tion and annealing temperature to
maximize desired product and
minimize “primer-dimers”.
Reaction working, but
unexpected extra prod-
ucts or smear is seen.
Nonspecific hybridization occurring
during setup.
Program a hot start into the
protocol.
Reaction component concentration
too high or too low.
Check concentrations of compo-
nents. May need to reoptimize
magnesium concentration.
Annealing temperature too low.
Reoptimize annealing temperature.
Protocol contains a wrong value.
Use List utility to check protocol’s
temperature control method, tem-
peratures, and times.
Template not of sufficient purity.
Check extraction and purification
protocols. Add additional purifica-
tion steps if necessary.
Multiple templates or host DNA in
sequencing reactions.
Check nucleic acid preparations
by gel electrophoresis.