Creating Programs
Tech Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com 5-25
•
Use the backspace key to delete the current value if it is not 03.
•
Enter 03 from the numeric keypad, and click once to accept the change.
•
Position the cursor (text pointer) in the time:second field of step 3 and
select by clicking once.
•
Use the backspace key to delete the current value if it is not 00.
•
Enter 00 from the numeric keypad, and click once to accept the change.
Entering a Goto Step
Step 4 of our program incorporates a goto step designed to cycle a portion of the program
a predetermined number of times. We have chosen to cycle back to step 2 and repeat
steps 2 and 3 an additional 29 times. A graphical program can only contain one goto step.
This goto step cannot be the first or only step in the protocol.
If there is a goto step (blue arrow) in the displayed protocol:
•
Select the goto step.
•
Select <Delete Step>.
To add a goto step to our protocol:
•
Select step 3 (the last step to be included in the goto loop).
•
Select <GoTo>.
A small red arrow will appear under step 3.
•
Select step 2 (the first step to be included in the goto loop).
A red arrow will now extend from the end of step 3 to the beginning of step 2,
and the default number of cycles the loop will execute, 25 X, will be displayed.
•
Select the number of cycles and delete using the backspace key.
•
Enter 29 from the numeric keypad, and click once to accept the change.
Our program will now run with 30 cycles.
Step 3
gradient step
Step 4
goto step