![BIO RAD PTC-0240 Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/bio-rad/ptc-0240/ptc-0240_operation-manual_2753175082.webp)
Managing and Editing Programs
Tech Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com
6-7
Inserting a Step
Now we will insert an annealing temperature step.
•
Highlight step 2.
Recall that new steps are inserted AFTER the highlighted step.
•
Select <Temp> from the advanced programming window.
The Temperature Step window appears. Select an incubation Temperature of
60°C, a hold Time of 3 minutes, and the Beep on Target option. Enter param-
eters as described in Chapter 5.
Your program should appear as follows.
Temperature Control Mode: Calculated
Lid Control Mode: Constant at 100°C
1. Incubate at 94°C for 00:01:00
Beep on Target
2. Incubate at 92°C for 00:00:30
Beep on Target
3. Incubate 60°C for 00:03:00
Beep on Target
4. Cycle to step 2 for 29 more times.
5. Incubate at 10°C forever
End
Editing a Step
Step 2 of the program ADV#1 includes a 30 second incubation. We wish to change
that incubation to 25 seconds.
•
Select Step 2 by double-clicking.
You are presented with the Temperature Step window. The Temperature field
shows 92°C, and the Time field shows 30 seconds.
•
Select the Time:Sec field.
•
Change the time from 30 to 25.
•
Select <OK>.