Unpacking and Moving the DNA Engine Tetrad 2 Cycler
Please read the “Unpacking Instructions for the DNA Engine Tetrad
®
2 Thermal Cycler”
insert in order to properly and safely remove the instrument from its packaging. Always
enlist the help of another individual when moving or lifting the DNA Engine Tetrad 2 cycler.
In order to lift the instrument, grasp the DNA Engine Tetrad 2 cycler from underneath the
chassis (placing your hands on either side of the DNA Engine Tetrad 2 cycler in between
the feet of the instrument) and lift. Do not insert the Alpha
™
units until the instrument is
located in its final place.
Packing Checklist
After unpacking the DNA Engine Tetrad 2 cycler, check to see that you have received the
following:
•
One DNA Engine Tetrad 2 base
•
Four spare fuses
•
One power cord
•
One external mouse device
•
DNA Engine Tetrad 2 Peltier Thermal Cycler Operations Manual (this
document)
•
Unpacking Instructions for the DNA Engine Tetrad
®
2 Thermal Cycler
If any of these components are missing or damaged, contact Bio-Rad Laboratories to
obtain a replacement. Please save the original packing materials in case you need to
return the DNA Engine Tetrad 2 cycler for service.
Setting Up the DNA Engine Tetrad 2 Cycler
The DNA Engine Tetrad 2 cycler requires only minimal assembly, plugging in the power
cord and mounting the Alpha units. Insert the power cord plug into its jack at the back of
the machine (see figure 2-3 for location of jack), then plug the cord into a 220 V electrical
outlet. With the machine switched off, mount the Alpha units (see the “Installing an Alpha unit”
section in Chapter 4).
Caution:
Do not insert or remove an Alpha unit with the DNA Engine Tetrad 2
cycler turned on; electrical arcing can result. Read the safety warning at the
beginning of this manual on page iv regarding electrical safety when inserting