VDL (E3.1) 06/2020
Page 95/206
Locked operation
Provided that the user administration has been activated by the assignment of passwords for the different
authorization types, the
controller operation
is first locked after turning on the unit, recognizable by the
closed lock icon in the header.
In the locked view the controller provides all display functions. No setting functions are available.
The setpoints are shaded (light grey) in normal display. Changing them by direct entry in the fixed value
operating mode is not possible. The functional icons for setpoint entry and program start in the footer are
without function.
After turning on the unit, user log-in is required to operate the controller (chap. 12.2).
Operation without user log-in / without password-protection
If the password function has been deactivated, after turning on the unit without user log-in there are those
controller functions available, which correspond to the highest authorization level without a password
protection. There is no lock icon in the header.
9.8
Loading
Without technical ventilation the chamber must not be loaded with material containing sol-
vents.
The chamber cannot be loaded when turned off / in standby mode .
Before loading the material to be loaded, note:
When loading the chamber also in the context of intended use, an explosive mixture may form in the
working space. Therefore, a safety area of at least. 1 m in front of the oven must be considered.
•
Ensure equipotential bonding. The accessible installation and operating surface of the equipment must
be conductive. This installation and operating surface must be connected with the vacuum drying oven
according to the grounding plan (Chap. 6.8).
•
The personal protective equipment (PPE) of the operating personnel must be implemented ESD pro-
tected. This applies also to gloves.
•
Activate the technical ventilation (extraction):
The operator shall provide active ventilation (technical ventilation according to country-specific regula-
tions (TRBS 2152 Part 2 for Germany) prior to commissioning of the chamber. It must include the en-
tire installation area of the vacuum drying oven. Loading must take place under technical ventilation in
the safety area in front of the oven.
Содержание vdl 115
Страница 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Страница 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Страница 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Страница 197: ...VDL E3 1 06 2020 Page 197 206 30 Certificates and declarations of conformity 30 1 EU Declaration of Conformity ...
Страница 198: ...VDL E3 1 06 2020 Page 198 206 ...
Страница 199: ...VDL E3 1 06 2020 Page 199 206 ...
Страница 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Страница 207: ......