Normal operation when used correctly: Even without taking into account the pressure control, which is
also effective in practice, there is no ignition source during the drying process. This also applies in the
event that the system is ventilated before the drying process finishes.
Fault: Incorrect loading with inadmissible solvent with a too low ignition temperature: Due to the vacu-
um, there is no ignition source during the drying process. The drying process could safely be ended. In
order to cancel the drying process, it is the chamber must first have cooled to room temperature be-
fore ventilation and opening (chap. 9.10.4).
•
Taking into account the volume dependence of the auto-ignition temperature
The surface temperature is not to exceed 80% of the auto-ignition temperature of the gas or fluid,
measured in °C.
Action: The safety margin of 20% of the auto-ignition temperature. is sufficiently large (EN1127-
1:2019).
•
Safety pressure monitoring
with heater release by pressure switch
There is no explosive atmosphere in a vacuum. The heater is only released starting with a pressure
threshold of 100 mbar.
There is thus always a vacuum
≤ 100mbar during drying.
The protective measure of ignition source monitoring provides redundant safety during the drying pro-
cess: There are no hot surfaces at a pressure > 100 mbar, and, when used correctly, after the heater
is released the hot surfaces are always below the auto-ignition temperature of permitted materials.
•
Sweeping the area for electrical equipment and the controller housing (triangle instrument box)
with compressed air
The area for electrical equipment and the controller housing
are protected against penetration of ex-
plosive atmosphere by a technical ventilation system (minimum overpressure: 25 Pa). The user must
monitor the pressure on the manometer on the chamber front, especially when loading and removing
the loading material This reliably prevents explosive atmosphere to reach electrical or electronic parts
of the VDL safety vacuum drying oven (protection against solvent-containing ambient air).
•
Heater shutdown in the event of component failure
In case of failure of the controller Pt100 temperature sensor or the heater Pt100 temperature sensor.
The heater is turned off. On the controller an alarm message is displayed.
•
Measures for lacking / incorrect maintenance / adjustment / sensor drift
Proper maintenance requires that the controller sensor is correctly adjusted. The safety temperature
limiter (TL) provides sufficient safety against heating sensor drift.
Regular adjustment intervals for the controller sensor (annually) are specified (chap. 26.3).
•
Prevention of electrostatic charges / grounding plan
All chamber components are at a common potential. The interior parts of the chamber are grounded.
The outer powder-coated housing plates have a layer thickness of 60 µm and are connected with the
earth potential via the protective conductor connection.
A detailed grounding plan for the VDL, the pump module, the pump and the installation and loading
area is provided, all components are prepared accordingly for grounding: Connections are available to
ground the device and accessories. Achieving equipotential bonding according to the grounding plan
of the manufacturer is mandatory, see operating manual chap. 6.8.
To create the connection between the vacuum drying oven and the pump, an antistatic PTFE hose is
used. All conductive parts have the same electrical potential. Also the door handle is conductive.
Содержание vdl 115
Страница 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Страница 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Страница 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Страница 197: ...VDL E3 1 06 2020 Page 197 206 30 Certificates and declarations of conformity 30 1 EU Declaration of Conformity ...
Страница 198: ...VDL E3 1 06 2020 Page 198 206 ...
Страница 199: ...VDL E3 1 06 2020 Page 199 206 ...
Страница 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Страница 207: ......