VDL (E3.1) 06/2020
Page 134/206
“Test temp. limiter” setting menu.
To activate the test alarm for the first TL, enter
“1” and press the
Confirm
icon..
Submenu “Various” with activated test alarm
“1”.
Press the
Confirm
icon and go back to Normal
display.
Normal display in state of alarm (sample val-
ues)
The alarm message “Test temperature limiter”
Screen header flashes in red color. if the
buzzer is enabled there is an audible alert.
Press the
Alarm
icon.
List of active alarms.
Press the
Reset alarm
icon.
To deactivate the test alarm, go back to the “Test temp. limiter” setting menu and set the value to “0” (test
alarm deactivated).
Then pull out the power plug, let the chamber cool down for 20 minutes and then plug it in again.
For the chambers with a single safety temperature limiter (TL), the test is completed here.
For the chambers with two safety temperature limiters (TL), carry out another test for the second TL. To
do this, enter “2” in the “Test temp. limiter” setting menu to activate the test alarm for the second TL and
press the
Confirm
icon. Then proceed as described above.
Содержание vdl 115
Страница 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Страница 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Страница 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Страница 197: ...VDL E3 1 06 2020 Page 197 206 30 Certificates and declarations of conformity 30 1 EU Declaration of Conformity ...
Страница 198: ...VDL E3 1 06 2020 Page 198 206 ...
Страница 199: ...VDL E3 1 06 2020 Page 199 206 ...
Страница 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Страница 207: ......