16.4 Activating / deactivating the audible alarm (buzzer)
Path:
Main menu > Settings > Chamber
“Chamber” submenu (example).
In the field “Audible alarm” select the desired setting
“off“ or “on” and press the
Confirm
icon.
16.5 Test alarm of the safety temperature limiter (TL)
The safety temperature limiter (TL) provides temperature monitoring with a fixed switching threshold. It is
an essential element of the manufacturer's safety plan to prevent ignition and explosions (chap. 3.1).
Therefore, an annual function check by the operator is recommended. To do this, you can use the con-
troller test routine described following.
For chambers equipped with a single safety temperature limiter (TL), this is tested with the setting “1”. For
chambers with 2 TB, the test is run twice, first with the setting “1” and then again with the setting “2”.
Chambers with one TL
Chambers with two TL
VDL 23
VDL 23-UL
VDL 115
VDL 56
VDL 56-UL
VDL 115-UL
Path:
Main menu > Settings > Various
Submenu “Various”.