VDL (E3.1) 06/2020
Page 67/206
Extracted vapors or the occasionally used inert gas can endanger health and / or have a corrosive effect
on the oven and pump.
WARNING
Danger of damage to health by release of extracted vapors.
Corrosion on the chamber and the pump. Damage to health.
Remove the extracted vapors and/or inert gas via a suitable conduit from the pump
module into e.g., a fume extractor facility.
Directly connect the conduit to the corresponding pump outlets.
You can connect a laboratory cooling system to the emission condenser of the vacuum pump
VP4.
To operate the chemical membrane pumps VP4 (MZ2C EX) please refer to the enclosed
pump manufacturer's operating manual.
Confirm that the vacuum source is designed for a gas inlet temperature corresponding to the
used drying temperature, or take appropriate measures to cool down the extracted vapor be-
fore its entry into the vacuum source.
The ATEX Directive 2014/34/EU compliant vacuum pump offered by BINDER is designed for a gas
inlet temperature of 40 °C /
104 °F
max.
Do NOT exceed this temperature. If the gas inlet temperature
is too high and then becomes even warmer by compression in the pump, the resulting temperature (of the
gas-solvent mixture inside the pump) could exceed the solvent’s temperature class and auto-ignition tem-
perature .
The maximum permissible ambient temperature of the vacuum pump supplied by BINDER is 40 °C /
104 °F
.
6.5.4 Note on the use of a flame arrester
A flame arrester is not mandatory for VDL applications (TRBS 2152-4:2012). A pump designed for Zone 1
is equipped with a temperature sensor on the bearing which, in the event of a fault, causes the pump to
shut down immediately.
In general: In case of a risk of ignition before or after the vacuum pump, the operator must provide suita-
ble flame arresters in accordance with EN 12874. The suitability (chemical resistance, sufficient flow and
safety against clogging) must be ensured before commissioning.
Содержание vdl 115
Страница 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Страница 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Страница 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Страница 197: ...VDL E3 1 06 2020 Page 197 206 30 Certificates and declarations of conformity 30 1 EU Declaration of Conformity ...
Страница 198: ...VDL E3 1 06 2020 Page 198 206 ...
Страница 199: ...VDL E3 1 06 2020 Page 199 206 ...
Страница 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Страница 207: ......