VDL (E3.1) 06/2020
Page 188/206
Chamber size
23
56
115
External dimensions
Chamber with optional pump module
Width, net
mm /
inch
523 /
20.59
637 /
25.10
742 /
29.21
Height, net
mm /
inch
1425 /
56.10
1542 /
60.71
1669 /
65.71
Depth, net
mm /
inch
491 /
19.33
540 /
21.26
660 /
25.98
Depth, gross (including controller, door handle,
connections, pressure regulator)
mm /
inch
632 /
24.88
680 /
26.77
801 /
31.54
Wall clearances
Wall clearance, rear (minimum)
mm /
inch
100 /
3.94
100 /
3.94
100 /
3.94
Wall clearance, side (minimum)
mm /
inch
70 / 2
.76
70 / 2
.76
70 / 2
.76
Internal dimensions
Width
mm /
inch
285 /
11.22
400 /
15.75
506 /
19.92
Height
mm /
inch
285 /
11.22
400 /
15.75
506 /
19.92
Depth
mm /
inch
295 /
11.61
343 /
13.50
460 /
18.11
Interior volume
l /
cu.ft.
24 /
0.85
55 /
1.94
119 /
4.20
Racks
Number of expansion racks (aluminum), series
2
2
2
Number of expansion racks (aluminum), max.
4
5
6
Distance between the racks
mm /
inch
53
/ 2.09
62 /
2.44
68 /
2.68
Usable space per rack (width x depth)
mm /
inch
234 x 280
9.21 x 11.02
349 x 320
13.74 x 12.60
455 x 440
17.91 x 17.32
Permissible load per rack
Kg /
lbs
20 /
44
20 /
44
20 /
44
Permissible total load
Kg /
lbs
50 /
110
60 /
132
70 /
154
Weight
Weight (empty)
Kg /
lbs
72 /
159
104 /
229
158 /
348
Temperature data
Temperature range
degrees above room
temperature
°C /
°F
9 /
16.2
9 /
16.2
9 /
16.2
up to
°C /
°F
110 /
230
110 /
230
110 /
230
Temperature fluctuation at 100 °C /
212 °F
± K
0,1
0,1
0.1
Temperature uniformity (variation)
at 100 °C /
212 °F
± K
1,0
1,5
2,9
Heating up time to 100 °C /
212 °F
Min
150
140
170
Vacuum data
Vacuum connection with small flange
DN mm /
inch
16
16
16
Measuring access port with small flange
DN mm /
inch
16
16
16
Inert gas connection with flow-limiter
Adapter with hose olive
∅
mm /
inch
8
8
8
Vacuum range (adjustable)
mbar /
inHg
10 to 1100
10 to 1100
10 to 1100
Leak rate
bar/h
/
inHg
/h
1x10
-2
/
0.295
1x10
-2
/
0.295
1x10
-2
/
0.295
Ex
Classification
Ex-Classification of the entire chamber
according to ATEX Directive 2014/34/EU
II 2/3/- G IIB T3 Gb/Gc/- X
Electrical data
(model versions VDL023-230V, VDL056-230V, VDL115-230V)
Protection type according to EN 60529 - MB2
controller housing
IP
20
20
20
Protection type, VDL general (type plate)
IP
20
20
20
Содержание vdl 115
Страница 192: ...VDL E3 1 06 2020 Page 192 206 28 6 Dimensions 28 6 1 VDL 23 Dimensions in mm ...
Страница 193: ...VDL E3 1 06 2020 Page 193 206 28 6 2 VDL 56 Dimensions in mm ...
Страница 194: ...VDL E3 1 06 2020 Page 194 206 28 6 3 VDL 115 Dimensions in mm ...
Страница 197: ...VDL E3 1 06 2020 Page 197 206 30 Certificates and declarations of conformity 30 1 EU Declaration of Conformity ...
Страница 198: ...VDL E3 1 06 2020 Page 198 206 ...
Страница 199: ...VDL E3 1 06 2020 Page 199 206 ...
Страница 200: ...VDL E3 1 06 2020 Page 200 206 31 Product registration ...
Страница 207: ......