MK / MKT (E5) 03/2019
page 70/158
12. Temperature safety devices
12.1 Over temperature protective device (class 1)
The chamber is equipped with an internal temperature safety device, class 1 acc. to DIN 12880:2007. It
serves to protect the chamber and prevents dangerous conditions caused by major defects.
If the actual temperature exceeds the nominal temperature by approx. 20 °C, the over temperature protec-
tive device permanently turns off the chamber. The user cannot restart the chamber again. The protective
cut-off device is located internally. Only a service specialist can replace it. Therefore, please contact an
authorized service provider or BINDER Service.
12.2 Overtemperature safety controller class 2
The chambers are regularly equipped with an electronic overtemperature safety controller (temperature
safety device class 2 according to DIN 12880:2007). The safety controller is functionally and electrically
independent of the temperature control system and serves to protect the chamber, its environment and
the contents from exceeding the maximum permissible temperature.
With option temperature safety device class 3.3 (chap. 12.3), the safety controller is
not
used.
In this case it must be set to the maximum temperature.
Please observe the DGUV guidelines 213-850 on safe working in laboratories (formerly BGI/GUV-I 850-0,
BGR/GUV-R 120 or ZH 1/119, issued by the employers’ liability insurance association) (for Germany).
When the entered safety controller set-point is reached, the overtemperature safety controller turns off the
heating, fan, and refrigeration system. This condition (state of alarm) is indicated visually and additionally
with an audible alert if the buzzer is enabled (chap. 11.4). The alarm persists until the chamber cools
down below the configured safety controller setpoint and the alarm is manually reset.
Check the setting regularly and adjust it following changes of the set-point or charge.
The safety controller only activates after the set-point has been reached once.
12.2.1 Safety controller modes
You can select between “Limit (absolute)” and “Offset (relative)” safety controller mode
•
Limit:
Absolute maximum permitted temperature value
This setting offers high safety as a defined temperature limit will not be exceeded. It is important to
adapt the safety controller set-point after each modification of the temperature set-point. Otherwise, the
limit could be too high to ensure efficient protection, or, in the opposite case, it could prevent the con-
troller from reaching an entered set-point outside the limit range.
•
Offset:
Maximum overtemperature above any active temperature set point. The maximum tempera-
ture changes internally and automatically with every set-point change.
This setting is recommended for program operation. It is important to check the safety controller set-
point and safety controller mode occasionally, as it does not offer a fix, independent limit temperature
value, which would never be exceeded.
Содержание MK 115
Страница 129: ...MK MKT E5 03 2019 page 129 158 Heating up time MK 115 Cooling down time MK 115 Temp C Time min Time min Temp C ...
Страница 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Страница 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Страница 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Страница 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Страница 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Страница 143: ...MK MKT E5 03 2019 page 143 158 25 Certificates and declarations of conformity 25 1 EU Declaration of Conformity for MK ...
Страница 144: ...MK MKT E5 03 2019 page 144 158 ...
Страница 145: ...MK MKT E5 03 2019 page 145 158 ...
Страница 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Страница 147: ...MK MKT E5 03 2019 page 147 158 ...
Страница 148: ...MK MKT E5 03 2019 page 148 158 ...
Страница 150: ...MK MKT E5 03 2019 page 150 158 ...
Страница 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Страница 159: ......