MK / MKT (E5) 03/2019
page 124/158
24.5 MKT (E4) technical data
Chamber size
115
240
720
Exterior dimensions
Width, gross (including 18 mm for 1 ac-
cess port (MKT 115, 240), 36 mm for 2
access ports (MKT 720), with plug)
mm /
inch
1000/
39.37
1135/
44.69
1615/
63.58
Height, gross (incl. castors)
mm /
inch
1725/
67.91
1940 /
76.38
2005 /
78.94
Depth, gross (incl. cable and door handle)
mm /
inch
915/
36.02
1000/
39.37
1230/
48.43
Depth, gross (incl. cable and door handle)
with optional compressed air dryer
mm /
inch
1085 /
42.72
1170 /
46.06
1400 /
55.12
Depth, gross (incl. cable and door handle)
with voltage and frequency changer
mm /
inch
1530 /
60.24
1615 /
36.58
1845 /
72.64
Wall clearance rear (minimum)
mm /
inch
300 /
11.81
300 /
11.81
300 /
11.81
Wall clearance rear with optional com-
pressed air dryer or to set up the voltage
and frequency changer (minimum)
mm /
inch
1000 / 39.37
1000 / 39.37
1000 / 39.37
Wall clearance sides (minimum)
mm /
inch
200 /
7.87
200 /
7.87
200 /
7.87
Window width
mm /
inch
288 /
11.34
508 /
19.99
508 /
19.99
Window height
mm /
inch
222 /
8.74
300 /
11.81
300 /
11.81
Doors
Number of doors
1
1
1
Interior dimensions
Width
mm /
inch
600 /
23.62
735 /
28.94
1200 /
47.24
Height
mm /
inch
480 /
18.90
700 /
27.56
1020 /
40.16
Depth
mm /
inch
400 /
15.75
443 /
17.44
600 /
23.62
Interior volume
l /
cu.ft.
115 /
4.06
228 /
8.05
734 /
25.92
Racks
Quantity of racks (regular)
1
1
1
Quantity of racks (max.)
4
6
11
Maximum load per rack
Kg /
lbs.
30 /
66
30 /
66
40 /
88
Maximum permitted total load
Kg /
lbs.
60 /
132
70 /
155
160 /
353
Weight
Weight (empty)
Kg /
lbs.
305 /
672
380 /
838
610 /
1345
Weight (empty)
with optional compressed air dryer
kg /
lbs.
320 /
705
395 /
871
625 /
1378
Temperature data
Temperature range
°C
°F
-70 to +180
-94 to 356
-70 to +180
-94 to 356
-70 to +180
-94 to 356
Temperature fluctuation
± K
0.1 to 0.6
0.1 to 0.4
0.1 to 0.5
Temperature uniformity (variation)
± K
0.2 to 1.8
0.1 to 1.0
0.3 to 2.0
Average heating up time
acc. to IEC 60068-3-5
K/min.
5.3
5,0
4,5
Average cooling down time
acc. to IEC 60068-3-5
K/min.
4.2
4.2
4.2
Max. heat compensation at 25 °C /
77°F
W
1800
3000
5500
Содержание MK 115
Страница 129: ...MK MKT E5 03 2019 page 129 158 Heating up time MK 115 Cooling down time MK 115 Temp C Time min Time min Temp C ...
Страница 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Страница 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Страница 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Страница 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Страница 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Страница 143: ...MK MKT E5 03 2019 page 143 158 25 Certificates and declarations of conformity 25 1 EU Declaration of Conformity for MK ...
Страница 144: ...MK MKT E5 03 2019 page 144 158 ...
Страница 145: ...MK MKT E5 03 2019 page 145 158 ...
Страница 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Страница 147: ...MK MKT E5 03 2019 page 147 158 ...
Страница 148: ...MK MKT E5 03 2019 page 148 158 ...
Страница 150: ...MK MKT E5 03 2019 page 150 158 ...
Страница 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Страница 159: ......