MK / MKT (E5) 03/2019
page 118/158
23. Troubleshooting
Fault description
Possible cause
Required measures
General
Chamber without function.
No power supply.
Check connection to power supply.
Wrong voltage.
Check power supply for correct volt-
age (chap. 4.4).
Main power switch (1) not
turned on
Turn on the main power switch (1).
MK/MKT 115, 240, 720:
Rear
power switch (12) not turned
on.
Turn on the rear power switch (12).
Chamber fuse has responded.
Check chamber fuse and replace it if
appropriate. If it responds again,
contact BINDER service.
Controller defective.
Contact BINDER service.
Nominal temperature exceed-
ed by 20 °C due to chamber
failure. Over temperature pro-
tective device (class 1) re-
sponds.
Heating
Chamber heating permanently,
set-point not maintained.
Semiconductor relay defective.
Contact BINDER service.
Pt100 sensor defective.
Controller defective.
Controller not adjusted.
Calibrate and adjust controller.
Chamber doesn’t heat up.
Heating element defective.
Contact BINDER service.
Semiconductor relay defective
Chamber doesn’t heat up when
turned on.
Safety controller responds.
Inner chamber temperature
has reached the safety control-
ler setpoint. Safety controller
set too low.
Acknowledge the alarm on the con-
troller. Check temperature setpoint
setting. If appropriate, select suitable
safety controller setpoint (chap.
12.2).
Safety controller (chap. 12.2)
defective.
Contact BINDER service.
Safety controller class 2 responds Limit temperature reached.
Acknowledge the alarm on the con-
troller. Disconnect the chamber from
the power supply and let it cool
down. Detect cause and remove it.
Start up the chamber and check
control functions. If appropriate,
select suitable limit value.
Over-/under temperature safety
device class 2 (option) responds.
Limit temperature reached.
Acknowledge the alarm on the con-
troller. Disconnect the chamber from
the power supply and let it cool
down. Detect cause and remove it.
Press button “RESET CL 2.0” (2).
Start up the chamber and check
control functions. If appropriate,
select suitable limit value.
Содержание MK 115
Страница 129: ...MK MKT E5 03 2019 page 129 158 Heating up time MK 115 Cooling down time MK 115 Temp C Time min Time min Temp C ...
Страница 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Страница 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Страница 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Страница 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Страница 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Страница 143: ...MK MKT E5 03 2019 page 143 158 25 Certificates and declarations of conformity 25 1 EU Declaration of Conformity for MK ...
Страница 144: ...MK MKT E5 03 2019 page 144 158 ...
Страница 145: ...MK MKT E5 03 2019 page 145 158 ...
Страница 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Страница 147: ...MK MKT E5 03 2019 page 147 158 ...
Страница 148: ...MK MKT E5 03 2019 page 148 158 ...
Страница 150: ...MK MKT E5 03 2019 page 150 158 ...
Страница 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Страница 159: ......