24.4 MK (E4) technical data
Chamber size
56
115
240
720
Exterior dimensions
Width, gross (including 18 mm for 1 ac-
cess port MK 56, 115, 240), 36 mm for 2
access ports (MK 720), with plug)
mm
inch
745
29.3
1000
39.37
1135
44.69
1615
63.58
Height, gross (incl. castors)
mm
inch
1450
57.1
1725
67.91
1715
67.52
2005
78.94
Depth, gross (incl. cable and door handle)
mm
inch
835
32.9
915
36.02
1000
39.37
1230
48.43
Depth, gross (incl. cable and door handle)
with optional compressed air dryer
mm
inch
985
38.8
1085
42.72
1170
46.06
1400
55.12
Depth, gross (incl. cable and door handle)
with voltage and frequency changer
mm
inch
--
1530 /
60.24
1615 /
36.58
1845 /
72.64
Wall clearance rear (minimum)
mm /
inch
300 /
11.8
300 /
11.81
300 /
11.81
300 /
11.81
Wall clearance rear with optional com-
pressed air dryer or to set up the voltage
and frequency changer (minimum)
mm
inch
1000
39.37
1000
39.37
1000
39.37
1000
39.37
Wall clearance sides (minimum)
mm /
inch
300 /
11.8
200 /
7.87
200 /
7.87
200 /
7.87
Window width
mm /
inch
200 /
7.9
288 /
11.34
508 /
19.99
508 /
19.99
Window height
mm /
inch
288 /
11.3
222 /
8.74
300 /
11.81
300 /
11.81
Doors
Number of doors
1
1
1
1
Interior dimensions
Width
mm
inch
400
15.7
600
23.62
735
28.94
1200
47.24
Height
mm
inch
420
16.5
480
18.90
700
27.56
1020
40.16
Depth
mm /
inch
318 /
12.5
400 /
15.75
443 /
17.44
600 /
23.62
Interior volume
l /
cu.ft.
60 /
2.1
115 /
4.06
228 /
8.05
734 /
25.92
Racks
Quantity of racks (regular)
1
1
1
1
Quantity of racks (max.)
4
4
6
11
Maximum load per rack
Kg /
lbs.
15 /
33
30 /
66
30 /
66
40 /
88
Maximum permitted total load
Kg /
lbs.
60 /
132
60 /
132
70 /
155
160 /
353
Weight
Weight (empty)
Kg /
lbs.
165 /
364
255 /
562
321 /
708
570 /
1257
Weight (empty) with optional compressed
air dryer
kg /
lbs.
***
275 /
606
355 /
783
585 /
1290
Temperature data
Temperature range
°C
°F
-40 to +180
-40 to 356
-40 to +180
-40 to 356
-40 to +180
-40 to 356
-40 to +180
-40 to 356
Temperature fluctuation
± K
0.1 to 0.5
0.1 to 0.5
0.1 to 0.5
0.1 to 0.5
Temperature uniformity (variation)
± K
0.5 to 1.5
0.1 to 2.0
0.1 to 1.2
0.3 to 2.0
Average heating up time acc. to IEC
60068-3-5
K/min.
5.0
5.3
5.0
4.0
Average cooling down time acc. to IEC
60068-3-5
K/min.
5.0
5.0
4.5
4.5
Heating up time from -40 °C to 180 °C
minutes
75
55
50
96
Cooling down time from 180 °C to -40 °C
minutes
90
90
110
100
Max. heat compensation at 25 °C /
77°F
W
2000
1500
2000
3500
Содержание MK 115
Страница 129: ...MK MKT E5 03 2019 page 129 158 Heating up time MK 115 Cooling down time MK 115 Temp C Time min Time min Temp C ...
Страница 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Страница 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Страница 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Страница 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Страница 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Страница 143: ...MK MKT E5 03 2019 page 143 158 25 Certificates and declarations of conformity 25 1 EU Declaration of Conformity for MK ...
Страница 144: ...MK MKT E5 03 2019 page 144 158 ...
Страница 145: ...MK MKT E5 03 2019 page 145 158 ...
Страница 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Страница 147: ...MK MKT E5 03 2019 page 147 158 ...
Страница 148: ...MK MKT E5 03 2019 page 148 158 ...
Страница 150: ...MK MKT E5 03 2019 page 150 158 ...
Страница 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Страница 159: ......