MK / MKT (E5) 03/2019
page 110/158
21. Maintenance, cleaning, and service
21.1 Maintenance intervals, service
DANGER
Electrical hazard.
Danger of death.
∅
The chamber must NOT become wet during operation or maintenance works.
∅
Do NOT remove the rear panel of the chamber.
Disconnect the chamber before conducting maintenance work. Turn off the main power
switch and pull the power plug.
General maintenance work must be conducted by licensed electricians or experts au-
thorized by BINDER.
Maintenance work at the refrigeration system must only be conducted by qualified per-
sonnel who underwent training in accordance with EN 13313:2010 (e.g. a refrigeration
technician with certified expert knowledge acc. to Regulation (EC) nº 303/2008. Follow
the national statutory regulations.
Ensure regular maintenance work is performed at least once a year and that the legal requirements are
met regarding the qualifications of service personnel, scope of testing and documentation. All work on the
refrigeration system (repairs, inspections) must be documented in a service log book (equipment records).
In the course of this annual maintenance, a leak test shall be carried out in accordance with Regulation
(EU) 517/2014 (Article 4 and Article 10 (1) (b)).
The warranty becomes void if maintenance work is conducted by non-authorized personnel.
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.
With an increased amount of dust in the ambient air, clean the condenser fan several times a year. We
recommend checking the fan grid (behind the left maintenance access flap) every week. In case of visible
dirt accumulation, disconnect the chamber and clean the fan grid by suction.
We recommend taking out a maintenance agreement. Please consult BINDER Service.
BINDER telephone hotline:
+49 (0) 7462 2005 555
BINDER fax hotline:
+49 (0) 7462 2005 93555
BINDER e-mail hotline:
BINDER service hotline USA:
+1 866 885 9794 or +1 631 224 4340 x3 (toll-free in the USA)
BINDER service hotline Asia Pacific:
+852 390 705 04 or +852 390 705 03
BINDER service hotline Russia and CIS
+7 495 988 15 16
BINDER Internet website
http://www.binder-world.com
BINDER address
BINDER GmbH, post office box 102, D-78502 Tuttlingen
International customers, please contact your local BINDER distributor.
Содержание MK 115
Страница 129: ...MK MKT E5 03 2019 page 129 158 Heating up time MK 115 Cooling down time MK 115 Temp C Time min Time min Temp C ...
Страница 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Страница 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Страница 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Страница 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Страница 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Страница 143: ...MK MKT E5 03 2019 page 143 158 25 Certificates and declarations of conformity 25 1 EU Declaration of Conformity for MK ...
Страница 144: ...MK MKT E5 03 2019 page 144 158 ...
Страница 145: ...MK MKT E5 03 2019 page 145 158 ...
Страница 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Страница 147: ...MK MKT E5 03 2019 page 147 158 ...
Страница 148: ...MK MKT E5 03 2019 page 148 158 ...
Страница 150: ...MK MKT E5 03 2019 page 150 158 ...
Страница 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Страница 159: ......