Chapter 11. Error Lists and Device States
78
MicroPolar 2 (++) LB 567
11.3
Measurement Error and Error Prompts
Code Error
Possible cause
50
Sigma of phase is
too large
The measured phase ex-
ceeds the allowed Sigma
limit.
54
No system calibra-
tion done
The system calibration has
not yet been carried out.
55
Insertion attenua-
tion is not reached.
See Software Manual, chap-
ter 8
60
Current in 1 out of
range
The enabled current input
has not yet been calibrated
or is not occupied.
61
Current in 2 out of
range
The enabled current input
has not yet been calibrated
or is not occupied.
62
PT100 temperature
out of range
The enabled PT100 input has
not yet been calibrated or is
not yet occupied.
70
Concentration out
of range
The concentration is outside
the process limits.
71
Concentration 2 out
of range
Concentration 2 is outside
the process limits.
72
Loading value 1 be-
low minimum load
Below the minimum load for
concentration 1
73
Loading value 2 be-
low minimum load
Below the minimum load for
concentration 1
74
Loading compensa-
tion inactive current
input
upper & lower limit
invalid
Current input is outside the
range.
90
Current output 1
out of range
The concentration calculated
on the basis of the current is
outside the current range
91
Current output 2
out of range
The concentration calculated
on the basis of the current is
outside the current range
120
Date/Time
not set
Please enter the date and
time.
For all other error messages, please contact the Berthold
Technologies Service.
Содержание LB 567
Страница 2: ......
Страница 3: ......
Страница 36: ......
Страница 60: ......
Страница 71: ...Chapter 7 Certificates MicroPolar 2 LB 567 71 Chapter 7 Certificates 7 1 EC Declaration of Conformity...
Страница 72: ...Chapter 7 Certificates 72 MicroPolar 2 LB 567 7 2 Frequency License...
Страница 74: ...Chapter 8 Technical Drawings 74 MicroPolar 2 LB 567 8 2 Electrical Wiring Diagram...
Страница 77: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 77 8 3 3 Type LB 3543 11 MP nominal width 50 mm adapter...
Страница 78: ...Chapter 8 Technical Drawings 78 MicroPolar 2 LB 567 8 3 4 Type LB 3547 11 MP nominal width 65 mm options...
Страница 80: ...Chapter 8 Technical Drawings 80 MicroPolar 2 LB 567 8 3 6 Type LB 3545 11 MP nominal width 80 mm...
Страница 82: ...Chapter 8 Technical Drawings 82 MicroPolar 2 LB 567 8 3 8 Type LB 3544 11 MP nominal width 100 mm options...
Страница 83: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 83 8 3 9 Type LB 3544 21 MP nominal width 100 mm DN100 PN16...
Страница 85: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 85 8 3 11 Type LB 3548 11 MP nominal width 150 mm...
Страница 88: ...Chapter 8 Technical Drawings 88 MicroPolar 2 LB 567...
Страница 89: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 89 8 4 2 Spiral antenna...
Страница 90: ...Chapter 8 Technical Drawings 90 MicroPolar 2 LB 567 8 5 Installation Proposal at the Conveyor Belt...
Страница 91: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 91 8 6 Installation Proposal at the Measuring Chute...
Страница 94: ...Notes 94 MicroPolar 2 LB 567 Notes...
Страница 95: ...Notes MicroPolar 2 LB 567 95 Notes...
Страница 96: ......
Страница 98: ......
Страница 102: ......
Страница 104: ......
Страница 106: ......
Страница 108: ......
Страница 136: ......
Страница 142: ......
Страница 156: ......
Страница 172: ......
Страница 176: ......
Страница 185: ...Notes MicroPolar 2 LB 567 89 Notes...
Страница 186: ...Notes 90 MicroPolar 2 LB 567 Notes...