3.8
Assembly on a Pipeline
The evaluation unit is installed close to the flow cell to keep the
HF cable between evaluation unit and probe as short as possible.
The shorter the cable connection, the better the stability of the
measurement. The standard length is 2 m and the maximum
length of the HF cable is 10 m.
The flow cell is integrated into the existing pipeline system or in
a bypass. The orientation of the flow cell can either be vertical or
horizontal. To avoid possible sedimentary deposits, vertical in-
stallation in a riser is preferred (see Figure 3-18).
The installation should preferably be close to a sample sampling
point to ensure representative sampling for calibration.
A representative temperature signal (current signal or PT100)
should be connected to the evaluation unit for product tempera-
ture compensation.
Figure 3-18:
Typical measurement
arrangement on
a pipeline
Pipeline inline
or in bypass
Product flow
Sampling
Reference line
with N-connector
HF-cable quad
Evaluation unit
Distance:
typical 2 m
Orientation:
HF-connections
facing down
Micro-Polar
Содержание LB 567
Страница 2: ......
Страница 3: ......
Страница 36: ......
Страница 60: ......
Страница 71: ...Chapter 7 Certificates MicroPolar 2 LB 567 71 Chapter 7 Certificates 7 1 EC Declaration of Conformity...
Страница 72: ...Chapter 7 Certificates 72 MicroPolar 2 LB 567 7 2 Frequency License...
Страница 74: ...Chapter 8 Technical Drawings 74 MicroPolar 2 LB 567 8 2 Electrical Wiring Diagram...
Страница 77: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 77 8 3 3 Type LB 3543 11 MP nominal width 50 mm adapter...
Страница 78: ...Chapter 8 Technical Drawings 78 MicroPolar 2 LB 567 8 3 4 Type LB 3547 11 MP nominal width 65 mm options...
Страница 80: ...Chapter 8 Technical Drawings 80 MicroPolar 2 LB 567 8 3 6 Type LB 3545 11 MP nominal width 80 mm...
Страница 82: ...Chapter 8 Technical Drawings 82 MicroPolar 2 LB 567 8 3 8 Type LB 3544 11 MP nominal width 100 mm options...
Страница 83: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 83 8 3 9 Type LB 3544 21 MP nominal width 100 mm DN100 PN16...
Страница 85: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 85 8 3 11 Type LB 3548 11 MP nominal width 150 mm...
Страница 88: ...Chapter 8 Technical Drawings 88 MicroPolar 2 LB 567...
Страница 89: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 89 8 4 2 Spiral antenna...
Страница 90: ...Chapter 8 Technical Drawings 90 MicroPolar 2 LB 567 8 5 Installation Proposal at the Conveyor Belt...
Страница 91: ...Chapter 8 Technical Drawings MicroPolar 2 LB 567 91 8 6 Installation Proposal at the Measuring Chute...
Страница 94: ...Notes 94 MicroPolar 2 LB 567 Notes...
Страница 95: ...Notes MicroPolar 2 LB 567 95 Notes...
Страница 96: ......
Страница 98: ......
Страница 102: ......
Страница 104: ......
Страница 106: ......
Страница 108: ......
Страница 136: ......
Страница 142: ......
Страница 156: ......
Страница 172: ......
Страница 176: ......
Страница 185: ...Notes MicroPolar 2 LB 567 89 Notes...
Страница 186: ...Notes 90 MicroPolar 2 LB 567 Notes...