54733-10BA2L
3 – 434
02.2017
8 Error Handling
Volume 3
102
Device data-set error
Parameter set invalid
If the error appears on restart after a
new software has been loaded, then you
have to do a
Factory Reset
. If the fault
is still present then, an additional
Software Reset
is required.
If the fault appears during operation,
then you can fix the error only by
replacing the detector or the detector
electronics.
105
Real time clock not valid
Date invalid
If the device was turned off, check or
update the date and time.
If the unit was in operation when the
error occurred, then there is a defect in
the hardware. The error can be elimi-
nated only by replacing the detector
electronics.
106
Test mode active
A safety parameter could
not be set.
Check or reset all test settings in the
Service menu.
107
Watchdog reset
A line failure has trig-
gered an error in the
program flow and the
device has been
restarted automatically.
If the error occurs again, the detector
electronics has to be replaced. If the
exchange is unsuccessful, search for
electromagnetic interference.
108
Safety Parameter invalid
Safety parameter invalid
Check in the Diagnostics menu which of
the safety parameters could not be set
and why.
320
HV voltage
A faulty measurement
was detected in the high
voltage.
The error can be eliminated only by
replacing the detector electronics or the
photomultiplier.
321
Generated HV voltage
322
HV reached its limit value
323
HV average is 20% lower than default HV
324
HV average is 40% higher than default HV
325
Lower PMT current limit is exceeded
The PMT current is
>50
μ
A.
Either there is strong radiation interfer-
ence, or the photomultiplier (PMT) is
defective and therefore has to be to
exchanged.
326
Upper PMT current limit is exceeded
The PMT current is
>100µA.
327
Temperature sensor deviation
The redundantly mea-
sured electronics tem-
peratures differ more
than 10° C from each
other.
The error can be eliminated only by
replacing the detector electronics.
328
Temperature warning
The temperature in the
detector has exceeded
75° C, or fell below -35°
C.
Assemble water cooling, respectively
assemble a trace heating system.
329
Temperature out of allowed limits
The temperature in the
detector has exceeded
80° C, or fell below
-40° C.
We recommend to return the detector
for repair to the manufacturer, even if
the detector appears to be functioning
still.
405
RS-485 Communication error
Master-slave communi-
cation disturbed.
Check the wiring between master and
slave, the detector address and the sup-
ply voltage.
406
Remote device warning
A slave returns a warn-
ing message.
Go to the Diagnostics menu and check
in the local slave table which warning
message is indicated for which slave.
Cod
e
Error message
Error reason
Error Handling
Содержание LB 480
Страница 2: ......
Страница 3: ...1 2 3 Volume 1 Safety Manual Volume 2 Installing SENSseries Volume 3 Operation with HART Communicator...
Страница 11: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 1 Safety Manual...
Страница 12: ...54733 10BA2L 1 12 02 2017 Volume 1...
Страница 22: ...54733 10BA2L 1 22 02 2017 4 Transport and Assembly Volume 1...
Страница 55: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 55 Volume 1 5 Explosion Protection 1 5 9 ATEX Certificate...
Страница 56: ...54733BA26 1 56 02 2017 5 Explosion Protection Volume 1...
Страница 57: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 57 Volume 1 5 Explosion Protection 1...
Страница 58: ...54733BA26 1 58 02 2017 5 Explosion Protection Volume 1...
Страница 59: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 59 Volume 1 5 Explosion Protection 1...
Страница 60: ...54733BA26 1 60 02 2017 5 Explosion Protection Volume 1...
Страница 61: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 61 Volume 1 5 Explosion Protection 1...
Страница 62: ...54733BA26 1 62 02 2017 5 Explosion Protection Volume 1...
Страница 63: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 63 Volume 1 5 Explosion Protection 1...
Страница 64: ...54733BA26 1 64 02 2017 5 Explosion Protection Volume 1...
Страница 66: ...54733BA26 1 66 02 2017 5 Explosion Protection Volume 1...
Страница 67: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 67 Volume 1 5 Explosion Protection 1...
Страница 68: ...54733BA26 1 68 02 2017 5 Explosion Protection Volume 1...
Страница 69: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 69 Volume 1 5 Explosion Protection 1...
Страница 70: ...54733BA26 1 70 02 2017 5 Explosion Protection Volume 1...
Страница 71: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 71 Volume 1 5 Explosion Protection 1...
Страница 72: ...54733BA26 1 72 02 2017 5 Explosion Protection Volume 1...
Страница 73: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 73 Volume 1 5 Explosion Protection 1...
Страница 108: ...54733 10BA2L 1 108 02 2017 7 Functional Safety Volume 1 7 13 Functional Safety Certificate...
Страница 109: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 109 Volume 1 7 Functional Safety 1...
Страница 110: ...54733 10BA2L 1 110 02 2017 7 Functional Safety Volume 1...
Страница 112: ...54733 10BA2L 1 112 02 2017 8 Visual Inspection Volume 1...
Страница 128: ...54733 10BA2L 1 128 02 2017 9 Radiation Protection Volume 1...
Страница 147: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 147 Volume 1 1 Notes...
Страница 149: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 2 Installing SENSseries...
Страница 150: ...54733 10BA2L 2 150 02 2017 Volume 2...
Страница 170: ...54733 10BA2L 2 170 02 2017 1 System Description Volume 2...
Страница 240: ...54733 10BA2L 2 240 02 2017 4 Repair Maintenance and Upkeep Volume 2...
Страница 254: ...54733 10BA2L 2 254 02 2017 5 Technical Information Volume 2 LB 480 Dimensions in mm Abmessungen in mm...
Страница 288: ...54733 10BA2L 2 288 02 2017 5 Technical Information Volume 2...
Страница 297: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 2 297 Volume 2 2 Notes...
Страница 299: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 3 Operation with HART Communicator...
Страница 300: ...54733 10BA2L 3 300 02 2017 Volume 3...
Страница 390: ...54733 10BA2L 3 390 02 2017 3 Getting Started via the HART Communicator Volume 3...
Страница 410: ...54733 10BA2L 3 410 02 2017 6 Functional Processes Volume 3...
Страница 428: ...54733 10BA2L 3 428 02 2017 7 Explanations Volume 3...
Страница 440: ...54733 10BA2L 3 440 02 2017 8 Error Handling Volume 3...
Страница 444: ...54733 10BA2L 3 444 02 2017 9 Setup Protocol Volume 3 Path Parameters Unit Standard SETUP...
Страница 445: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 3 445 Volume 3 3 Notes...
Страница 449: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 449 Volume 1 Notes...