Chapter 2: Operation
5
STEP-BY-STEP SUMMARY OF OPERATING PROCEDURES
SET-UP
INSPECT THE SYSTEM
Power cord
Reservoir Support
Disposable Set
Large Reservoir and holder, if needed
Inspect the system to ensure that you have all
necessary components.
Use only supplied power cord.
IV POLE MOUNTING
IV Pole: 5 wheel, maximum diameter 1
1/4"
Install the Support Assembly 30" from
the ground, if not already installed.
Mount the FMS2000 on the IV Pole
above the Support Assembly
Install the Reservoir Support app. 9"
above the top of the system
CAUTION:
Check that the system is securely clamped
to an IV pole and will not tip over
1. Install the support assembly (support clamp and
washer) approximately 30" from the ground.
While holding clamp closed, loosen the
screw to open up the clamp. Install clamp
on the IV pole, holding clamp close and
tighten screw using the supplied 3/16 Allen
wrench.
Snap the plastic washer onto the IV pole
above the support clamp.
2.
Lift up on the “Pole Clamp Release Handle” to
open. Mount the system onto the IV pole, above
the support assembly, by pushing down on the
pole clamp release handle. Check that the
system is locked in place before proceeding.
3. Clamp the reservoir support onto the IV pole
approximately 9" above the FMS2000.
Make certain that there is nothing
obstructing the air vents at the bottom of
the system.